B-303 Evaluating an Automated Urinalysis System: Analytical Performance and Verification Strategies
Sergio Tufik, Victor Dorsa, Pamila Cunha, Rosa Maria Silva, Graziele Ferreira, Jussimara Nurmberger, Carolina Piccinini, Mariana Viana
Key points
From the paper's conclusion- This study aimed to verify the performance of the UN2000 system and to describe the analytical strategies adopted for method verification.
Abstract
Background Urinalysis is among the most frequently requested laboratory tests in healthcare services, as multiple biochemical and cellular parameters can be assessed, supporting clinical decision, diagnosis, and therapeutic monitoring. Therefore, automated urinalysis systems are widely used and essential in high-throughput laboratory routines, representing valuable tools to ensure result safety, speed, and reliability. In this context, pre-implementation verification is required to demonstrate test suitability and analytical safety. This study aimed to verify the performance of the UN2000 system and to describe the analytical strategies adopted for method verification. Methods The UN2000 (Sysmex) integrates urine biochemical analysis using reagent strips (UC-3500) with particle counting by flow cytometry (UF-5000). Due to distinct analyte profiles, different strategies were applied for sample and data processing. Urine biochemistry precision was assessed using two levels of controls, processed 25 consecutive times to evaluate repeatability. Particle counting precision was measured in quintuplicate, over five days, to evaluate repeatability and reproducibility using two control levels. Acceptance criteria were the manufacturer’s standard deviation (SD). For comparison, 20 urine samples were analyzed using the UN2000 and the manual reference method (gold standard). Quantitative comparison was performed for pH, specific gravity, erythrocytes, and leukocytes using allowable total error (TEa) based on biological variation from the European Federation of Clinical Chemistry and Laboratory Medicine or analytical quality control criteria as acceptance limits. Qualitative analysis was applied to the remaining strip parameters and other particles using a semi-quantitative grading system, with acceptance criteria defined as a kappa index >=75% and approximately 100% agreement. Linearity and carryover for erythrocytes and leukocytes were evaluated by flow cytometry according to the manufacturer’s instructions. Linearity was assessed using six dilution samples in duplicate, with acceptance of r² >0.99. For carryover, 21 samples were processed in a predefined sequence, with acceptance defined as carryover below the calculated error limit. Data were analyzed using EP Evaluator software (version 12.4). Results Precision results of both reagent strip and flow cytometry analyses were within the accepted SD limits at the 95% confidence level. All qualitative comparative analyses demonstrated satisfactory performance, with kappa values >=75%, agreement of approximately 100%, and correlations >=95%. Quantitative comparative data were in accordance with the acceptance criteria, with pH and specific gravity differences between methods remaining within TEa in 100% of cases. Although correlation coefficients were <0.99, this result was expected due to intrinsic methodological differences and did not affect the performance of the UN2000 nor impair clinical interpretation. Linearity and carryover assessments passed, with r² values of 1.0 and carryover limits within three SDs of consecutive low-concentration aliquots, respectively, indicating excellent proportionality across the measurement range. Conclusion Verification of the system demonstrated satisfactory performance according to the established criteria. When comparing automated urinalysis systems with manual methods, intrinsic methodological differences must be considered, and analytical strategies should be adapted to the characteristics of each. These findings highlight the importance of verifying automated urinalysis systems, with the potential to improve laboratory efficiency and emphasize the relevance of appropriate analytical strategies to ensure reliable result interpretation.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- B-001 Limitations of Urine Dipstick Testing and the Clinical Value of Creatinine-Normalized Quantitative Urinalysis in Urologic and Renal Disease DetectionClinical Chemistry · 2026
- A-271 Standardization of a Single Sterile Urine Collection Tube: Diagnostic Stewardship Applied to the Pre-Analytical PhaseClinical Chemistry · 2026
- Water bath prewarming of urine samples to facilitate automated microscopy and urinalysis.Laboratory medicine · 2026
- Automated monitoring of urine output in hospitalized patients with indwelling urinary catheters: a clinical evaluation on the cardiology wardNetherlands Heart Journal · 2026
- B-302 Body Fluids as Complex Analytical Metrices: Challenges in Validation and Analytical Performance of Automated Biochemistry AnalyzersClinical Chemistry · 2026
- Automated urinalysis for predicting urine culture positivity and presumptive Gram classification: a retrospective study at a cancer hospital.Frontiers in cellular and infection microbiology · 2026