A-218 Ensuring Deionized Water Availability for Clinical Analyzers: A Clinical Engineering Strategy for Operational Continuity in a Laboratory Network
Cyra Araujo, Ronaldo Silva, Patricia Costa, Jonas Oliveira, Rafael Jacomo, Lidia Nery, Vitor Domingos, João Oliveira,
Key points
From the paper's conclusion- This study describes a Clinical Engineering strategy designed to ensure continuous deionized water availability for 27 systems supporting laboratory analyzers across 12 Technical Operational Centers.
Abstract
Background Treated/deionized water systems are critical infrastructures for the operation of automated clinical analyzers. Continuous water quality and availability directly affect analytical performance, reagent stability, internal system integrity, and ultimately the reliability of laboratory results. Interruptions in water supply or degradation of physicochemical parameters may lead to immediate analyzer shutdown, test repetition, delayed result reporting, and potential clinical risk. In geographically distributed laboratory networks, variability in water supply quality — influenced by regional characteristics — makes standardized consumable management a significant technical challenge. This study describes a Clinical Engineering strategy designed to ensure continuous deionized water availability for 27 systems supporting laboratory analyzers across 12 Technical Operational Centers. Methods An individualized technical assessment was conducted for each water system, considering: historical analyzer downtime associated with water system failures; identification of critical consumables (deionization cartridges, filters, UV lamps, and filtration elements); observed practical lifespan of each component; manufacturer recommendations; and local incoming water characteristics (conductivity, organic load, and particulates), which vary by region. Clinical Engineering correlated historical consumption data with real operational parameters, adjusting consumable sizing according to the specific conditions of each system. The analysis recognized that practical consumable lifespan may be shorter than the manufacturer’s nominal expectation, particularly in regions with greater variability in the water supply quality. A preventive planning model was implemented with strategic inventory and scheduled replenishment, prioritizing elimination of stock-out risk for critical consumables. Monitored indicators included: downtime events due to lack of consumables, stability of water quality parameters (conductivity, resistivity, and Total Organic Carbon (TOC), and impact on analyzer availability. Results During the analyzed period, none of the 27 systems experienced operational interruption due to lack of consumables. Consequently, no clinical analyzer downtime was attributed to deionized water unavailability. The individualized approach improved predictability of component replacement and reduced technical alarms related to water quality deviations. Stability of physicochemical parameters mitigated analytical failure risk and minimized test repetition. Although the primary objective was operational continuity, the model also generated an estimated financial optimization of approximately USD 116,000 per year. These savings resulted from precise technical planning, elimination of emergency purchases, and avoidance of premature component replacement. More significant than the economic impact was the consolidation of a stable operational environment, ensuring uninterrupted support for diagnostic processes. Conclusion Ensuring continuous deionized water availability requires an individualized technical approach based on real operational data, regional variability of water supply quality, and critical analysis of consumables. The structured intervention by Clinical Engineering eliminated interruptions related to water systems, sustaining uninterrupted analyzer operation. In laboratory medicine, where operational continuity is directly linked to patient safety, technical management of water systems must be recognized as a strategic component essential to analytical reliability and clinical stability.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Sustainability comparison of three clinical laboratory analytical systems in Spain.Clinical chemistry and laboratory medicine · 2026
- Functionality status and drivers of non-functionality of the standardised laboratory analysers in Tanzania mainland: Findings from a National Equipment Harmonisation and Standardisation Mapping ExercisemedRxiv · 2026
- Global variability in dental unit waterline safety standards: bridging the gap between guidelines and practice.Frontiers in oral health · 2026
- B-281 Comparison of Analytical Error Frequencies Between Closed-System and Open-System Clinical Chemistry AnalyzersClinical Chemistry · 2026
- An Integrated Evaluation of Drinking Water Quality and Environmental Planning Implications Using WQI and Exceedance Analysis: Al-Doura Water Treatment Plant, BaghdadJournal of Engineering · 2026
- Multi-source drinking water systems and disinfection strategies: evidence from inspections, MadridEuropean Journal of Public Health · 2026