Quantitative Fetal Hemoglobin Measurement: GazelleTM Point-of-Care Electrophoresis versus Standard HPLC Methodology
T. N. Williams, R. Connon, R. Muhindo, D. Amorut, P. Ongodia, C. Okalebo, J. Abeso, F. Alaroker,
Key points
From the paper's conclusion- Mean (standard deviation) baseline HbF values by HPLC and Gazelle (N=2042) were 11.4% (SD 5.9%) versus 16.3% (SD 7.5%), respectively, while mean on-treatment HbF values (N=1013) were 18.9% (SD 9.0%) versus 25.4% (SD 11.5%), respectively.
Abstract
Quantitative measurement of fetal hemoglobin (HbF) using high-performance liquid chromatography (HPLC) is standard but expensive and widely unavailable in low-income settings. The newer Gazelle method, which employs single-use alkaline electrophoresis cartridges, provides a portable and relatively inexpensive potential alternative. In the current study, we tested 3055 venous blood samples collected from 1780 children with sickle cell anemia who were enrolled into the H-PRIME trial by both Gazelle and the Bio-Rad D-10 HPLC instrument, before and during hydroxyurea treatment. Data reinterpretation was required to quantify accurately high HbF values derived using the D-10. Mean (standard deviation) baseline HbF values by HPLC and Gazelle (N=2042) were 11.4% (SD 5.9%) versus 16.3% (SD 7.5%), respectively, while mean on-treatment HbF values (N=1013) were 18.9% (SD 9.0%) versus 25.4% (SD 11.5%), respectively. While overall agreement was good (Lins concordance 0.76, 95% CI 0.75-0.78), HbF values derived using Gazelle were modestly higher than HPLC values (Bland-Altman agreement mean +5.4%, 95%CI -2.4% to +13.2%), with the difference between measures increasing at higher HbF percentages. With strong concordance and more easily interpretable HbF values, Gazelle represents an attractive and inexpensive alternative to HPLC for HbF quantification in the low-income settings where most patients with sickle cell anemia live.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Validation of the Gazelle point-of-care device for hemoglobin variant analysis in a predominantly pediatric population: performance comparison with conventional electrophoresis.Frontiers in medicine · 2026
- Validation of HemoCue method and development of a revised hemoglobin cut-off to detect anemia in children aged 6–24-monthsPLoS ONE · 2026
- Evaluation of Different Methods of Haemoglobin Estimation Among Malaria-Infected Patients.Journal of clinical laboratory analysis · 2026
- Whole blood hemolysis detection on the GEM® Premier™ 7000: performance across hemoglobin A, S, and F.Clinical chemistry and laboratory medicine · 2026
- Capillary Collection Kit Can Facilitate Newborn Screening for Hemoglobinopathies: Utilisation of Existing Infrastructure in Resource Constrained Setting.Indian journal of hematology & blood transfusion : an official journal of Indian Society of Hematology and Blood Transfusion · 2026
- Unexpectedly low labile HbA1c levels may serve as a laboratory clue for HbC and HbE variants in routine HPLC HbA1c testingRevista Romana de Medicina de Laborator · 2026