Capillary Collection Kit Can Facilitate Newborn Screening for Hemoglobinopathies: Utilisation of Existing Infrastructure in Resource Constrained Setting.
Richa Juneja, M Vasanthakumari, Akriti Khare, Nishant Banait, Rasika Gadkari
Department of Pathology, All India Institute of Medical Sciences, MIHAN, Nagpur, 441108 Maharashtra India. · Department of Neonatology, All India Institute of Medical Sciences, MIHAN, Nagpur, 441108 Maharashtra India.
Key points
From the paper's conclusion- We evaluated diagnostic accuracy of heel prick samples collected in capillary collection kits from BIO-RAD in comparison with cord blood samples for NBS for hemoglobinopathies on Bio-Rad Variant II analyzer.
Abstract
Hemoglobinopathies like sickle cell disorder (SCD) and thalassemia are common in India. Newborn screening (NBS) in high-prevalence areas is crucial to initiate early intervention. Traditionally NBS is conducted using dried blood spots (DBS) on special high performance liquid chromatography (HPLC) or capillary zone electrophoresis (CZE) analyzer, however, HPLC analyzers for venous blood samples are widely available. We evaluated diagnostic accuracy of heel prick samples collected in capillary collection kits from BIO-RAD in comparison with cord blood samples for NBS for hemoglobinopathies on Bio-Rad Variant II analyzer. Cord blood and heel prick samples were collected from 100 newborns at a tertiary care center in Central India. Both samples were subjected to HPLC on Bio-Rad variant II analyzer. Sensitivity and specificity of capillary kit sample to detect hemoglobinopathy was compared with cord blood. Quantification of fetal Hb (Hb F), adult Hb (Hb A) and variant Hb was compared across 2 samples using bland altman plot. Capillary kit sample showed sensitivity of 93% and specificity of 99% in detection of hemoglobinopathies. Statistical analysis suggested excellent interclass correlation between both samples through bland Altman plot for quantification of various types of hemoglobin. Heel prick sample collected using Bio-Rad Capillary collection kit has > 90% sensitivity and specificity to detect hemoglobinopathy in newborns. Thus, it can be used to offer NBS in remote areas of developing countries on routine HPLC analyzer bypassing the need of cord blood samples and dedicated costly analyzers to process DBS samples.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Agreement Between Venous and Capillary Haemoglobin Concentrations in a High-Burden Urban Indian Setting: A Cross-Sectional Analysis.Diagnostics (Basel, Switzerland) · 2026
- Validation of the Gazelle point-of-care device for hemoglobin variant analysis in a predominantly pediatric population: performance comparison with conventional electrophoresis.Frontiers in medicine · 2026
- Antenatal Screening for Haemoglobinopathies and Maternal Outcomes at a Rural Healthcare Facility in Piparia, Gujarat, India: A Prospective Cohort StudyDOAJ (DOAJ: Directory of Open Access Journals) · 2026
- The Efficiency of Capillarys 2 Flex Piercing, Premier Resolution-HPLC, and VARIANT II-HPLC for Hemoglobin Constant Spring Detection.Hemoglobin · 2026
- Coverage of antenatal haemoglobinopathy screening in the Capital Region of Denmark: a laboratory-based evaluation of screening implementation.Scandinavian journal of clinical and laboratory investigation · 2026
- Haemoglobinopathies in Newborn Screening: Epidemiological Findings and Exploratory Analysis of HbA0 Distribution in a Population-Based Neonatal Cohort from Northern Italy.International journal of neonatal screening · 2026