Validation of HemoCue method and development of a revised hemoglobin cut-off to detect anemia in children aged 6–24-months
Mustafa Mahfuz, Ar-Rafi Khan, S. M. Tafsir Hasan, Md. Shabab Hossain, A. H. M. Rezwan, Mehjabin Tishan Mahfuz, Md. Ashraful Alam, Tahmeed Ahmed
International Centre for Diarrhoeal Disease Research
Key points
From the paper's conclusion- This study highlights substantial variability between HemoCue and DCM hemoglobin measurements, which prevented development of a reliable correction factor. The proposed HemoCue cut-off of 11.0 g/dL may improve anemia detection in children under two years of age when compared with the currently recommended threshold.
Abstract
BACKGROUND: Accurate hemoglobin (Hb) measurement is essential for determining anemia prevalence at the population level. The HemoCue method, commonly used in resource poor settings, has been shown to overestimate Hb levels compared to the direct cyanmethemoglobin (DCM) method, leading to an underestimation of anemia. We aimed to validate the HemoCue method against the DCM method and establish an optimal cut-off value for diagnosing anemia in children under two. METHODS: Data were collected from children under two years of age as part of the Etiology, Risk Factors, and Interactions of Enteric Infections and Malnutrition and the Consequences for Child Health and Development study in Bangladesh, conducted in Mirpur, Dhaka, from February 2010 to February 2017. Using HemoCue 201 instrument and DCM method, Hb levels were measured in 846 venous blood samples collected from 589 children. Quadratic regression was explored to derive a correction factor; however, substantial dispersion in HemoCue measurements prevented establishment of a reliable correction model. Youden's Index was used to identify the optimal cut‑off. While it provided a strong threshold, we prioritized achieving at least 70% sensitivity and specificity to confirm balanced diagnostic performance. RESULTS: The median (IQR) Hb concentrations estimated by HemoCue and DCM were 11.6 (10.5, 12.5) g/dL and 10.8 (10.1, 11.5) g/dL, respectively. Using a cut-off of <10.5 g/dL for anemia, the prevalence was significantly higher with the DCM method (35.2% vs. 23.2%, p < 0.001). Considering DCM as the gold standard, the sensitivity and specificity of the HemoCue method were 93% and 53%, respectively. In the test/train analysis, the newly identified cut-off of 11.00 g/dL, indicated an anemia prevalence of 34.2% in the testing dataset, with a sensitivity of 84% and specificity of 69%. CONCLUSION: This study highlights substantial variability between HemoCue and DCM hemoglobin measurements, which prevented development of a reliable correction factor. The proposed HemoCue cut-off of 11.0 g/dL may improve anemia detection in children under two years of age when compared with the currently recommended threshold.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Agreement Between Venous and Capillary Haemoglobin Concentrations in a High-Burden Urban Indian Setting: A Cross-Sectional Analysis.Diagnostics (Basel, Switzerland) · 2026
- Comparative Diagnostic Accuracy of Clinical Pallor at Different Anatomical Sites for Detecting Anemia in Pregnant Women Using HemoCue® as the Reference Standard.Cureus · 2026
- Evaluation of Different Methods of Haemoglobin Estimation Among Malaria-Infected Patients.Journal of clinical laboratory analysis · 2026
- Validation of Non-Invasive Hemoglobin Measurement and Its Utility for Anemia Screening in Pregnancy: A Cross-Sectional Diagnostic Accuracy Study.Diagnostics (Basel, Switzerland) · 2026
- Quantitative Fetal Hemoglobin Measurement: GazelleTM Point-of-Care Electrophoresis versus Standard HPLC MethodologymedRxiv · 2026
- Accuracy of minimally invasive hemoglobin monitoring in the perioperative period: transcutaneous hemoglobin monitoring and point-of-care testing devices: prospective observational study.Journal of clinical monitoring and computing · 2026