Evaluation of Different Methods of Haemoglobin Estimation Among Malaria-Infected Patients.
Mina Ofosu, Jessica Ntiamoah, Patricia Sam Barko, Prince Elikem Torli
Department of Laboratory Technology, Faculty of Health Sciences, Kumasi Technical University, Kumasi, Ghana.
Key points
From the paper's conclusion- The three methods demonstrated good agreement and broadly comparable Hb measurements. The findings support the practical use of portable Hb meters where point-of-care testing is required, while recognising that agreement does not establish diagnostic accuracy or analytical validation.
Abstract
BACKGROUND: Haemoglobin (Hb) estimation is essential for assessing anaemia, particularly in malaria-endemic settings. Automated Full Blood Count (FBC) analysers, cyanmethemoglobin spectrophotometry, and portable Hb meters are commonly used, but differences in analytical performance may affect measurements. METHODS: This cross-sectional study compared Hb measurements obtained using an automated FBC analyser, the cyanmethemoglobin method, and a portable Hb meter among 100 malaria-infected patients aged 16-65 years at Kumasi South Hospital, Ghana. Measurements were compared using Pearson's correlation coefficient, intraclass correlation coefficient (ICC), Lin's concordance correlation coefficient (CCC), Bland-Altman analysis, percentage error, and Passing-Bablok regression. RESULTS: The three methods showed very strong correlations (r = 0.955-0.992, p < 0.001), excellent ICC (single measures = 0.969; average measures = 0.990), and high concordance (CCC = 0.95-0.99). Mean biases were -0.06 g/dL for FBC analyser versus portable Hb meter, -0.25 g/dL for FBC analyser versus cyanmethemoglobin, and -0.20 g/dL for portable Hb meter versus cyanmethemoglobin. The corresponding 95% limits of agreement were -0.59 to 0.47, -1.48 to 0.98, and -1.35 to 0.95 g/dL, respectively. Passing-Bablok slopes were 1.00 for all comparisons. The FBC analyser and portable Hb meter showed the closest agreement. CONCLUSION: The three methods demonstrated good agreement and broadly comparable Hb measurements. The findings support the practical use of portable Hb meters where point-of-care testing is required, while recognising that agreement does not establish diagnostic accuracy or analytical validation.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Validation of Non-Invasive Hemoglobin Measurement and Its Utility for Anemia Screening in Pregnancy: A Cross-Sectional Diagnostic Accuracy Study.Diagnostics (Basel, Switzerland) · 2026
- Validation of HemoCue method and development of a revised hemoglobin cut-off to detect anemia in children aged 6–24-monthsPLoS ONE · 2026
- A-342 Method Comparison of Blood Gas Total Hemoglobin and Automated CBC Hemoglobin Measurements in a Neonatal Intensive Care UnitClinical Chemistry · 2026
- Agreement Between Venous and Capillary Haemoglobin Concentrations in a High-Burden Urban Indian Setting: A Cross-Sectional Analysis.Diagnostics (Basel, Switzerland) · 2026
- Quantitative Fetal Hemoglobin Measurement: GazelleTM Point-of-Care Electrophoresis versus Standard HPLC MethodologymedRxiv · 2026
- B-233 Haemoglobin Variants and Their Impact on HbA1c Measurement: Experience with Two Routine MethodsClinical Chemistry · 2026