Whole blood hemolysis detection on the GEM® Premier™ 7000: performance across hemoglobin A, S, and F.
Antoine Puravet, Benjamin Bouthors, Benjamin Rieu, Quentin Montade, Bruno Pereira, Marina Brailova, Vincent Sapin, Damien Bouvier
Biochemistry and Molecular Genetics Department, Centre Hospitalier Universitaire Clermont-Ferrand, Clermont-Ferrand, France. · Department of Perioperative Medicine, Centre Hospitalier Universitaire Clermont-Ferrand, Clermont-Ferrand, France.
Key points
From the paper's conclusion- The GEM ® Premier™ 7000 incorporates an onboard hemolysis detection system based on multi-wavelength co-oximetry, allowing direct estimation of free hemoglobin in whole blood samples.
Abstract
OBJECTIVES: Point-of-care blood gas analyzers are essential in critical care settings but remain vulnerable to pre-analytical interferences, particularly hemolysis, which may lead to erroneous potassium results. The GEM® Premier™ 7000 incorporates an onboard hemolysis detection system based on multi-wavelength co-oximetry, allowing direct estimation of free hemoglobin in whole blood samples. This study evaluated the analytical performance of this system across different hemoglobin variants, including fetal hemoglobin (HbF) and hemoglobin S (HbS). METHODS: Fifty patient samples undergoing routine arterial blood gas and biochemical testing were analyzed using the GEM® Premier™ 7000, and Atellica Solution analyzers. Experimental hemolysis was induced to evaluate the performance of the onboard hemolysis detection system. In a second step, hemoglobin overload experiments using HbA0, HbF, and HbS were performed in samples from 10 healthy volunteers to assess the robustness of hemolysis detection across different hemoglobin variants. RESULTS: Experimental hemolysis significantly increased the hemolysis index and potassium concentrations measured on the GEM® Premier™ 7000, with median hemolysis index values rising from 1 to 6 and median potassium concentrations from 3.6 to 9.8 mmol/L (p<0.001). The GEM® Premier™ 7000 accurately detected increasing concentrations of free HbA0, HbS, and HbF throughout the overload experiments. CONCLUSIONS: The integrated hemolysis detection system reliably detected clinically significant hemolysis in whole blood samples, supporting more reliable point-of-care blood gas testing.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- A-409 Comparative analytical performance of cartridge-based blood gas analyzers and the clinical impact of integrated whole-blood hemolysis detectionClinical Chemistry · 2026
- Hemolysis in neonatal whole-blood gas testing: prevalence and association with potassium.European journal of pediatrics · 2026
- A-367 Assessment of Hemolysis Interference on HbA1c Measurement Using the Siemens Healthineers DCA Vantage AnalyzerClinical Chemistry · 2026
- A-357 Characterization of Flea-Mixing-Induced Hemolysis in Blood-Gas Capillary TubesClinical Chemistry · 2026
- The Efficiency of Capillarys 2 Flex Piercing, Premier Resolution-HPLC, and VARIANT II-HPLC for Hemoglobin Constant Spring Detection.Hemoglobin · 2026
- A-215 Performance Evaluation for a New Generation of G6PD Assay(G6PDH-pro) With Whole-blood Without Pre-centrifugation on Mindray BS-600M AnalyzerClinical Chemistry · 2026