Radiance-Invariance Evaluation of a Compact LED-Based Source for Imager Radiance Transfer.
Alberto J Ruiz, Edwin A Robledo, Eammon A Littler
Key points
From the paper's conclusion- Conclusions: The presented methods provide an approach for evaluating compact LED-based sources for invariant-radiance emission and suitability for SI-traceable radiance transfer.
Abstract
SIGNIFICANCE: Fluorescence imaging remains largely qualitative and device-specific, limiting reproducibility and intersystem comparisons. SI-traceable radiometric imager characterization requires calibrated sources suitable for transferring known radiance to imaging systems. AIM: Establish and evaluate a compact, calibrated solid-state radiometric emitter target (RET) for SI-traceable radiance transfer, and assess methods for evaluating radiance invariance, imager responsivity, and extension of radiance transfer to digitally defined regions of interest (ROIs). APPROACH: The radiance invariance of an LED-based RET was evaluated by measuring its radiance across varying source-to-aperture distances and collection apertures. A previously described radiance-transfer formalism based on calibrated source radiance and collection geometry was implemented to determine imager responsivity ( R λ ), which was evaluated across varying working distances and lens apertures. Grubbs' and MAD outlier tests and one-way ANOVA were used to assess radiance and responsivity invariance. Finally, responsivity obtained using digitally defined ROIs was compared with matched physical apertures to assess ROI-based radiance transfer. RESULTS: The measured RET radiance remained invariant across the tested source-to-aperture distances and collection apertures, with no significant differences observed across distance (p = 0.88) or aperture (p = 0.15). Imager responsivity remained stable across working distances for both tested lens configurations (p = 0.48 and p > 0.99). Responsivity was also invariant across lens apertures, except for deviations at f/11 that were consistent with practical aperture tolerances. Digitally defined ROIs reproduced responsivity values obtained using matched physical apertures, demonstrating that the radiance-transfer approach can be extended to image-defined regions. CONCLUSIONS: The presented methods provide an approach for evaluating compact LED-based sources for invariant-radiance emission and suitability for SI-traceable radiance transfer. The characterized RET supported stable imager responsivity across changes in imaging geometry and extension to digitally defined ROIs. These methods can be adapted to other LED-based sources, with further angular characterization and uncertainty analysis needed to establish broader applicability.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Investigation of the active laser pyrometer with reflection for platinum dynamic surface true temperature measurement.Optics letters · 2026
- Calibration of digital cameras for remote measurements of luminance and illuminance of object surfacesJournal of Sensors and Sensor Systems · 2026
- Numerical Evaluation of Sparse Residual Correction for Joint Luminance and Chromaticity Calibration in Mini-LED DisplaysElectronics · 2026
- Active imaging radiometry—not always 1/R²Applied Optics · 2026
- Traceable state-matched radiometric calibration for gated line-scan spectroscopy at 100 kHz.Optics express · 2026
- Optical Scintillation Imaging for Dynamic Source Tracking Dosimetry During Prostate HDR BrachytherapyarXiv · 2026