Calibration of digital cameras for remote measurements of luminance and illuminance of object surfaces
Valentyn Sudomyr, Yaroslav Osadtsa
Key points
From the paper's conclusion- The developed technology makes it possible to transform the common camera into an effective two-dimensional measuring tool for remote measurements of the object luminance and illuminance.
Abstract
A method for calibrating commercial digital cameras for the remote measurement of the luminance and illuminance of object surfaces is presented in this paper. Traditional photometric methods based on sequential point measurements by luminance meters and lux meters are labour intensive, whereas the proposed approach makes it possible to create two-dimensional maps of the distribution of lighting parameters within the entire frame very quickly. The method is based on the application of unprocessed raw image data to ensure the linearity of the image sensor response and to avoid nonlinear transformations performed by the camera's built-in processor. The processing included conversion of the data into CIE 1931 X Y Z colour space, where the Y coordinate directly corresponds to the pixel luminance. A Canon EOS 1000D camera was used in this investigation. Calibration of this camera was carried out using semiconductor light sources with three colour temperatures (from 3053 to 7198 K) within the luminance range from 3.5 to 21.6 cd m −2 . It is determined that in order to ensure the required accuracy, separate calibration models (in the form of the dependence of the Y coordinate on exposure) should be applied for each value of the aperture number. The resulting models demonstrated high approximation quality ( R 2 >0.994). Verification on a diffuse-reflective surface showed an average deviation of 10.47 % for a sample of 21 525 control areas for first experimental dataset and an average deviation of 9.48 % for a sample of 2277 control areas for a second experimental dataset. Meanwhile, within the optimal range of the normalised signal ( Y =0.1–0.4), the accuracy was higher when using models for each aperture value. The developed technology makes it possible to transform the common camera into an effective two-dimensional measuring tool for remote measurements of the object luminance and illuminance.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Smartphones as Portable Tools for Reliable Color Determination of Metal Coatings Using a Colorimetric Calibration CardCoatings · 2025
- Luminance measurement model for high-dynamic-range imaging systemsMetrologia · 2026
- Boundary-Continuous Spectral Demoiréing for Pixel-Level Luminance Measurement of Display PanelsInformation · 2026
- Comparative Analysis of Image Quality Assessment Between Digital Signal-to-Noise Ratio (SNR) Camera and Smartphone Camera for Dental Photography.Cureus · 2026
- Spectral Optimization of Programmable LED Illumination for Smartphone-Based Colorimetric Biosensor Readout.Biosensors · 2026
- Standardising ocular surface photography for artificial intelligence: current limitations and a proposed roadmap.Progress in retinal and eye research · 2026