Optics letters

Investigation of the active laser pyrometer with reflection for platinum dynamic surface true temperature measurement.

Xinda Zhang, Ningtao Wang, He Wang, Yunlong Zhao, Baolin An, Xiuli Li, Ang Huang, Zundong Yuan,

National Institute of Metrology, Beijing 100029, China. · Tsinghua University, Beijing 100084, China.

Key points

From the paper's conclusion
  • Therefore, in this study, based on the ALPR principle, we propose a temperature measurement model that incorporates a time‑dependent term, and we build a reflective dynamic measurement system.

Abstract

The dynamic temperature evolution in semiconductor manufacturing has been difficult to measure accurately by non‑contact means, largely due to the effect of emissivity variations-where unknown and changing emissivity is one of the primary reasons. To address this issue, active laser pyrometry with reflection (ALPR) offers an emissivity‑independent technical approach. However, existing studies have mainly focused on steady‑state temperature measurements, without in‑depth analysis of the measurement response when the target temperature changes dynamically, and no public reports have been made on dynamic true‑temperature measurements. Therefore, in this study, based on the ALPR principle, we propose a temperature measurement model that incorporates a time‑dependent term, and we build a reflective dynamic measurement system. Experimental investigations are conducted on the dynamic true temperature of a platinum surface in the range of 1020 K to 1140 K. Within the temperature interval covered in this paper, the spectral radiation exhibits good linearity. The measurement results are compared with the thermocouple-extrapolated temperatures at corresponding time points, yielding a relative deviation of 1.0%. This work provides robust technical support for traceability of dynamic surface temperatures.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Investigation of the active laser pyrometer with reflection for platinum dynamic surface true temperature measurement — Science Skim