Traceable state-matched radiometric calibration for gated line-scan spectroscopy at 100 kHz.
Xuan-Sheng Huang, Shu-Bin Liu, Tian-Shuai Yuan, Chun-Bo Zou, Xiao-Qiang Lu
College of Physics and Information Engineering, Fuzhou University, Fuzhou 350108, China.
Key points
From the paper's conclusion- We present a traceable state-matched framework that treats the assembled optical path and synchronized acquisition state as the calibration object.
Abstract
Conventional radiometric calibration assumes stable illumination and an invariant system response, whereas microsecond-gated weak-light spectroscopy is highly sensitive to synchronized gating and readout. We present a traceable state-matched framework that treats the assembled optical path and synchronized acquisition state as the calibration object. A certified integrating sphere anchors the 1-kHz response, while frozen source-level, gate-width, and gain transfers are validated on held-out states. At 100 kHz, the system operated with a nominal 10 μs row interval and a programmed 1 μs electrical gate setting. Because the high-speed response was 22.96% lower than at 1 kHz, qualified scalar direct anchors were established at 550 and 600 nm. Held-out errors ranged from --1.56 to +0.40 formula units, and direct-anchor expanded uncertainties were 5.28% and 5.15% (k = 2). Plasma measurements demonstrate calibration-linked apparent spectral radiance retrieval from weak, nonrepeatable transient events.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Continuously tunable attenuation stage for weak-light radiance calibration using an integrating sphere and a dual-aperture tube.Applied optics · 2026
- Ultra-high precision speckle spectrometer enabling radio-frequency scale resolution of atomic spectraNature Communications · 2026
- Dynamic calibration method for the system constant of high-spectral-resolution lidar with airborne experimental verification.Optics express · 2026
- Absolute Calibration of a Broadband Transmission Grating Spectrometer from 5 to 400 nm with Uncertainty EstimatesarXiv · 2026
- Spectral flattening system for astronomical calibration sources based on a digital micromirror device.Optics express · 2026
- Dual-band scanned-wavelength infrared laser-induced fluorescence of CO with time-domain modeling for 1 kHz thermometry.Optics express · 2026