Penggunaan Reagen Modifikasi pada Pemeriksaan Hemoglobin Metode Sianmethemoglobin
Muhammad Ahyat Ridhoni, Rusmini Rusmini, Linda Yuliana
Poltekkes Kemenkes Banjarmasin
Key points
From the paper's conclusion- All data were normally distributed (p > 0.05) with mean hemoglobin levels closely approximating the Drabkin reagent across all control levels.
Abstract
Cyanmethemoglobin method hemoglobin level measurement is still widely used as a reference method due to its good accuracy, but the relatively expensive cost of standard reagents drives the need for a more economical alternative reagent. This laboratory experimental study with a Post-test Only Control Group Design aimed to determine the difference in hemoglobin level measurement results using a modified surfactant-based reagent—formulated from 5 ml of commercial dishwashing liquid, 10 ml of analytical grade Coco Glucoside, and distilled water up to 1000 ml—compared to commercial Drabkin reagent. Measurements were performed on low, normal, and high level commercial control blood with 16 repetitions each. Data analysis included the Shapiro–Wilk normality test, paired sample t-test, and calculation of bias percentage. All data were normally distributed (p > 0.05) with mean hemoglobin levels closely approximating the Drabkin reagent across all control levels. Paired sample t-test results indicated no significant differences between the two reagents at low (p = 0.719), normal (p = 0.084), and high levels (p = 0.632), with bias percentages of −0.20%, −0.47%, and −0.06%, respectively. Consequently, the modified surfactant-based reagent produced analytical performance comparable to Drabkin reagent and has the potential to be developed as an economical alternative following further validation.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- A-204 Chemical modification of the hemolysis index with methylene blue on an automated chemistry analyzer: The holy grail?Clinical Chemistry · 2026
- Análise comparativa de dois reagentes diferentes de glicose de marcas Mindray e Biosystem, mensurados no analisador bioquímico Mindray BS – 120Revista fisio&terapia. · 2026
- Evaluation of Different Methods of Haemoglobin Estimation Among Malaria-Infected Patients.Journal of clinical laboratory analysis · 2026
- Validation of HemoCue method and development of a revised hemoglobin cut-off to detect anemia in children aged 6–24-monthsPLoS ONE · 2026
- Multicenter evaluation of analytical performance and onboard stability of Mindray 10 concentrated biochemical reagentsJournal of Laboratory Medicine · 2026
- Novel method for measuring serum free hemoglobin in patients with hemolysisScientific Reports · 2026