A-204 Chemical modification of the hemolysis index with methylene blue on an automated chemistry analyzer: The holy grail?
Marisa Germaine, Joesph Wiencek
Key points
From the paper's conclusion- Interestingly, the lipemic index increased with higher concentrations of MB (range 154-30,495).
Abstract
Background Hemolysis is the leading cause of preanalytical specimen rejections. This phenomenon causes erroneous changes in concentrations of potassium and other substances released from damaged erythrocytes. The red color from free hemoglobin can also cause significant spectral interference in specific laboratory methods. To prevent reporting inaccurate results, automated chemistry analyzers can measure the hemolysis index (H-Index) to approximate free hemoglobin present. To date, there are resources available on how to mitigate hemolysis during collection, transport, and storage. However, there are no reported methods for removing free hemoglobin or modifying the H-index without impacting analyte recoveries. Methylene blue (MB) is a therapeutic blue thiazine dye that can interact with hemoglobin and influence its redox state, which may alter the H-Index. This study investigated the impact of MB on the H-Index measured on an automated chemistry analyzer. Methods Residual plasma from lithium heparin plasma separator tubes were collected. This pool was spiked with a prepared hemolysate to induce a measurable H-Index. Using this pool, two separate pools were then created to spike varying MB concentrations (range 0.001-0.5 mg/mL) or controls (water) according to the Clinical and Laboratory Standards Institute (CLSI) EP07 guidance document. The matrix of all spiking materials was kept below 5% as recommended by CLSI. All samples were measured in triplicate on a Roche cobas 503 for hemolysis, icterus and lipemia indices and basic metabolic panel (BMP). Acceptable total allowable error limits (TAE) for BMP analytes were defined by various available sources (e.g., CLIA, CAP). Results The average H-Index of the initial pool without and with hemolysate were determined to be 8 and 601, respectively. At the initial 0.001 mg/mL concentration of MB, the average H-Index decreased to 530 compared to the control (584) and further decreased to undetectable at 0.055 mg/mL (Figure 1). All analytes recovered from the BMP were within acceptable TAEs for the 0.055 mg/mL concentration of MB and control (data not shown). Interestingly, the lipemic index increased with higher concentrations of MB (range 154-30,495). The icterus index was not impacted at any MB concentrations assessed. Conclusion This is the first study to demonstrate the plausibility of chemically modifying the H-Index.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- B-346 Storage Stability of HbA1c in Whole Blood: Method-Dependent Interference on Ion-Exchange HPLC and Degradation Kinetics Characterized by LC–MSClinical Chemistry · 2026
- Quantifying phenylhydrazine-induced Heinz body formation in normal human erythrocytes by flow cytometry versus microscopy or spectrophotometry.Biology methods & protocols · 2026
- Methemoglobin Causes Collapsed WDF Scattergrams and a Low-SFL Mature RBC Population in RET Scattergrams on the XR-Series Automated Hematology Analyzer.International journal of laboratory hematology · 2026
- Novel method for measuring serum free hemoglobin in patients with hemolysisScientific Reports · 2026
- A-320 Verification of hemolysis levels to improve laboratory workflowClinical Chemistry · 2026
- Preanalytical stability of hemoglobin H inclusion body detection using rapid brilliant cresyl blue staining: a genotype-stratified stability study in EDTA blood stored at 4 °C.Laboratory medicine · 2026