Análise comparativa de dois reagentes diferentes de glicose de marcas Mindray e Biosystem, mensurados no analisador bioquímico Mindray BS – 120
João Segunda Cassoma Manuel, Pedro Nekundi Salgado, Anselmo Caquarta Victor, Amélia Katila Manuel da Rocha, Teresa Nambele Paulo Dumbo, Alfredo Francisco Golungo, Rui Walter Borges Aragão, Francisco Armindo Quiquelo,
Key points
From the paper's conclusion- Our study showed a strong correlation in the results of the two reagents, thus showing that both reagents can be reliable for use in the Mindray BS 120 biochemical analyzer and provide credible results without any doubt.
Abstract
The present study aimed to comparatively analyze the measurement of two different glucose reagents from the brands Mindray and Biosystem, processed on the Mindray BS 120 biochemical analyzer. This study is characterized as a cross-sectional study. The chosen setting was the Provincial Hospital of Cuanza Sul. The population studied consisted of all patients admitted to the Provincial Hospital of Cuanza Sul. The sample, however, was made up of 71 patients. The data obtained were as follows: the male gender accounted for 61% (n=43) of the patients tested, compared to the female gender, which accounted for 39% (n=28). patients who were tested. The age group with the highest frequency was 18 to 30 years at 28.2% (n=20), the average glucose levels of the two reagents showing a difference of 0.79 mg/dl, and a standard deviation of 1.8 mg/dL. As for the correlation variable, the two reagents also show strong correlations r = 0.991714247 ≈ 1. Our study showed a strong correlation in the results of the two reagents, thus showing that both reagents can be reliable for use in the Mindray BS 120 biochemical analyzer and provide credible results without any doubt.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Multicenter evaluation of analytical performance and onboard stability of Mindray 10 concentrated biochemical reagentsJournal of Laboratory Medicine · 2026
- B-030 Analytical Performance Evaluation of the Maccura Glycosylated Hemoglobin Analyzer G 02 (High-Performance Liquid Chromatography)Clinical Chemistry · 2026
- Blood Collection Tubes Utilized for Fasting Blood Sugar MeasurementPhilippine Journal of Pathology · 2026
- B-281 Comparison of Analytical Error Frequencies Between Closed-System and Open-System Clinical Chemistry AnalyzersClinical Chemistry · 2026
- A Portable Electrical Response Verification System for Blood Glucose Meters Based on Dynamic Microcurrent Reproduction.Sensors (Basel, Switzerland) · 2026
- Homeostasis de la Lipoconveniencia. Identificación de la Lipogénesis de Novo NO Enzimática mediante CapilaroscopíaImpact Research Journal · 2025