Independent Three-dimensional Dose Calculation from Log Files for Breast VMAT: A Measurement-free Quality Assurance Approach.
Rosa Angelica Petit Sevilla, Oscar Abel Apaza Blanco, Nayron Medeiros Soares, Maximiliano Musso, Ezequiel Agustín Cabrera Lacey, Luis C Bohorquez, Lucy Wolfsberger, Maria José Almada,
Department of Medical Physics, Instituto Zunino, Córdoba, Argentina. · Marie Curie Foundation, Zunino Institute, Córdoba, Argentina.
Key points
From the paper's conclusion- Independent 3D dose calculation is a reliable PSQA approach for breast VMAT, showing higher sensitivity than conventional 2D methods.
Abstract
PURPOSE: The purpose of this study was to evaluate and validate a measurement-free patient-specific quality assurance (PSQA) approach for volumetric modulated arc therapy (VMAT) breast cancer treatments based on independent three-dimensional (3D) dose calculation from Digital Imaging and Communications in Medicine (DICOM) plans and linear accelerator log files. MATERIALS AND METHODS: Seventy clinical breast VMAT plans (36 left-sided and 34 right-sided) were retrospectively analyzed. Independent 3D dose calculations were performed using RadCalc v7.3 based on the DICOM treatment plan (independent dose calculation [IDC] RTPlan) and on delivery log files acquired during the first treatment fraction (IDC LOGs). Results were compared with portal dosimetry (two-dimensional [2D]) and fluence-based analysis. Gamma passing rates (GPRs) were evaluated using 3%/2 mm and 2%/2 mm criteria under action (90%) and tolerance (95%) limits. Receiver operating characteristic (ROC) analysis was performed to evaluate the discriminatory capability. Sensitivity to delivery errors was investigated by introducing controlled multileaf collimator (MLC) shifts (0.5-2.0 mm) and collimator rotation errors (1°-3°). RESULTS: All PSQA methods met the TG-218 action limit (GPR ≥ 90%) at 3%/2 mm. Under the stricter 2%/2 mm criterion, statistically significant differences among quality assurance methods were observed (P < 0.001). ROC analysis demonstrated superior discriminatory performance for 3D dose-based methods, with area under the curve values of 0.9 for both IDC RTPlan and IDC LOGs, compared with 0.7 for portal dosimetry and 0.5 for fluence-based analysis. Error sensitivity analysis showed that stringent gamma criteria (1%/2 mm and 1%/1 mm) improved detection of subtle MLC and collimator deviations, with log file-based 3D reconstruction outperforming 2D approaches. CONCLUSIONS: Independent 3D dose calculation is a reliable PSQA approach for breast VMAT, showing higher sensitivity than conventional 2D methods.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Beyond Gamma Analysis: An Accessibility-oriented Framework for Comparative Evaluation of Patient-specific Radiotherapy Quality Assurance.Journal of medical physics · 2026
- Evidence-based design of a patient-specific quality assurance program through retrospective analysis of pretreatment and in vivo results.Physics and imaging in radiation oncology · 2026
- ESTRO-EPTN guidelines for independent dose calculation and log file-based patient-specific quality assurance in scanned particle therapy.Physics and imaging in radiation oncology · 2026
- A pathway away from measurement‐based patient‐specific QA: A review of contemporary quality assurance for external‐beam radiotherapy sub‐systemsJournal of Applied Clinical Medical Physics · 2025
- Predicting VMAT patient-specific quality assurance outcome using a machine learning workflow.Physical and engineering sciences in medicine · 2026
- Sensitivity in error detection of an electronic portal imaging device for patient-specific QA in head and neck VMAT plans.Radiological physics and technology · 2026