ESTRO-EPTN guidelines for independent dose calculation and log file-based patient-specific quality assurance in scanned particle therapy.
Loïc Grevillot, Oliver Jäkel, Abdallah Qubala, Julian Horn, Adam H Aitkenhead, Xiaoying Liang, Giuseppe Magro, Klará Badraoui-Čuprová,
MedAustron Ion Therapy Center, Wiener Neustadt, Austria. · Heidelberg Ion Beam Therapy Center (HIT), Department of Radiation Oncology, Heidelberg University Hospital, Heidelberg, Germany.
Key points
From the paper's conclusion- These recommendations aim to promote consistent and safe adoption across PT centers, while preserving the need for machine QA.
Abstract
BACKGROUND AND PURPOSE: Independent dose calculation (IDC) and log file-based patient-specific quality assurance (PSQA) are increasingly used in scanned particle therapy (PT) to address pitfalls of phantom-based verifications. Highly modulated treatment fields, steep dose gradients, and growing interest for adaptive treatments require quality assurance (QA) approaches that are cost effective, patient-specific, and closely linked to the delivered treatment. Practical guidance is therefore needed for the safe clinical implementation of IDC- and log file-based PSQA. MATERIALS AND METHODS: The ESTRO-EPTN (European Particle Therapy Network) developed consensus-based recommendations on IDC- and log file-based PSQA in PT. Current clinical practice, commissioning strategies, risk-management approaches, and available technical solutions were evaluated. Following a review of the literature, expert-based guidelines were developed covering clinical scopes, acceptance, commissioning and validation, integration with machine QA, data management, governance, and staged implementation. RESULTS: IDC- and log file-based PSQA were identified as complementary elements within a quality-management framework. Log file-based approaches enable evaluation of the delivered treatment fraction(s), while IDC provides computational verification of the planned dose. When appropriately validated, these methods increase sensitivity to potentially clinically relevant deviations and can substantially reduce reliance on phantom-based measurements. Safe application requires robust machine QA, characterization of model limitations, risk analysis, and data traceability. Recommendations applicable to all centers were developed for IDC. Log file-based PSQA recommendations include minimum requirements for the vendors. CONCLUSIONS: These recommendations aim to promote consistent and safe adoption across PT centers, while preserving the need for machine QA.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Monte Carlo simulations as the cornerstone of next-generation patient-specific quality assurance in particle therapy.Frontiers in oncology · 2026
- Independent Three-dimensional Dose Calculation from Log Files for Breast VMAT: A Measurement-free Quality Assurance Approach.Journal of medical physics · 2026
- Evidence-based design of a patient-specific quality assurance program through retrospective analysis of pretreatment and in vivo results.Physics and imaging in radiation oncology · 2026
- Five years of clinical application of independent Monte Carlo-based patient-specific quality assurance at Maastro Proton Therapy Center.Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB) · 2026
- Beyond Gamma Analysis: An Accessibility-oriented Framework for Comparative Evaluation of Patient-specific Radiotherapy Quality Assurance.Journal of medical physics · 2026
- A pathway away from measurement‐based patient‐specific QA: A review of contemporary quality assurance for external‐beam radiotherapy sub‐systemsJournal of Applied Clinical Medical Physics · 2025