Prolonged Rewarming Was Associated with Reduced Alkaline Phosphatase Bias in External Quality Assessment Materials: A Laboratory Study.
Dayun Xie, Xue Wen, Baoxiu Chen, Yali Li, Man Zhu, Wenbiao Zhou, Jiao Liu
Department of Laboratory Medicine, Chuxiong Yi Autonomous Prefecture People's Hospital. · Department of Plastic and Burn Surgery, West China Hospital, Sichuan University.
Key points
From the paper's conclusion- Recurrent negative alkaline phosphatase (ALP) bias in external quality assessment (EQA) materials despite stable internal quality control prompted investigation of handling conditions.
Abstract
Recurrent negative alkaline phosphatase (ALP) bias in external quality assessment (EQA) materials despite stable internal quality control prompted investigation of handling conditions. This study examined elapsed time after removal from frozen storage in five independent EQA materials. Materials were separately aliquoted (200 µL) and stored at -20 °C before testing. Different aliquots were analyzed on system Y at 30, 40, 50, and 60 min after removal, following routine shaking at 40 rpm for the first 20 min. Environmental records documented ambient temperatures of 20-25 °C; specimen temperatures were not measured. ALP and six comparison enzymes were measured in technical duplicate; 30-min ALP means were submitted for the first 2026 EQA round. Relative bias was calculated against method- or peer-group consensus targets, with materials as independent statistical units. Mean ALP bias (± standard deviation) changed from -6.7% ± 1.1% at 30 min to -1.7% ± 1.0% at 50 min and -1.3% ± 1.2% at 60 min. The exploratory 30-to-50-min comparison showed less negative bias in all five materials: mean paired change, 4.9 percentage points (t-based 95% confidence interval, 4.3-5.6; paired t-test: t = 21.248, df = 4, p < 0.001; two-sided exact sign test: p = 0.063). Comparison enzymes provided descriptive context without formal between-enzyme testing. Two subsequent EQA rounds yielded mean biases of 1.6% ± 1.2% and -0.2% ± 0.7%; different materials precluded paired comparisons or intervention validation. Longer elapsed time was associated with reduced negative ALP bias in these previously frozen materials. No matched refrigerated aliquots were tested, and commutability with patient specimens was not assessed. Measurement order remained confounded with elapsed time. Findings do not establish the original deviation's cause, improved analytical trueness, or an optimal rewarming interval. They support investigation of material-specific EQA handling; extrapolation to patient specimens requires separate evidence.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Discrepancies in plasma alkaline phosphatase reruns on Cobas c503 Roche analysers: a retrospective analysis and impact of dilution.Diagnosis (Berlin, Germany) · 2026
- B-032 On-board Stability and Multi-Manufacturer Comparison of Maccura Enhanced Alkaline Phosphatase AssayClinical Chemistry · 2026
- Effects of storage temperature and delayed analysis on routine biochemical and haematological test resultsScientific Reports · 2026
- Intact Parathyroid Hormone Reference Intervals: Importance of Pre-Processing for Indirect Reference Interval Estimation.Diagnostics (Basel, Switzerland) · 2026
- The P5P matrix paradox: Opposite AST method differences in quality-assessment materials and patient sera.Clinica chimica acta; international journal of clinical chemistry · 2026
- Short and Long Term Stability of Biochemical Analytes in Cattle Serum Under Varying Storage Conditions.Veterinary medicine and science · 2026