The P5P matrix paradox: Opposite AST method differences in quality-assessment materials and patient sera.

Po-Chih Chen, Hui-Ru Yang, Hsin-Chieh Lin, Yi-Hung Sung, Shyh-Jong Wu, Liang-Yin Ke

Department of Laboratory Medicine, Kaohsiung Medical University Chung-Ho Memorial Hospital, Kaohsiung Medical University, Kaohsiung 807377, Taiwan; Department of Medical Laboratory Science and Biotechnology, College of Health Sciences, Kaohsiung Medical University, Kaohsiung 807378, Taiwan. · Department of Laboratory Medicine, Kaohsiung Medical University Chung-Ho Memorial Hospital, Kaohsiung Medical University, Kaohsiung 807377, Taiwan.

Key points

From the paper's conclusion
  • This study compared P5P-free AST2 with P5P-containing activated aspartate aminotransferase (A-AST) using 2025 College of American Pathologists (CAP) proficiency-testing materials, Technopath quality-control data, and 137 patient sera analyzed on Abbott Alinity c systems.

Abstract

Previous studies have highlighted the importance of pyridoxal-5'-phosphate (P5P) supplementation in aminotransferase measurement. However, whether method differences are consistent across quality-assessment materials and patient sera remains unclear. This study compared P5P-free AST2 with P5P-containing activated aspartate aminotransferase (A-AST) using 2025 College of American Pathologists (CAP) proficiency-testing materials, Technopath quality-control data, and 137 patient sera analyzed on Abbott Alinity c systems. Analyses included percentage differences, Bland-Altman analysis, regression, and reference interval classification. A-AST results were lower than AST2 results in CAP and Technopath materials, with mean percentage differences of -5.9% and - 15.8%, respectively. In contrast, patient results were 14.7% higher with A-AST (95% CI, 13.0%-16.5%; p < 0.001). The mean paired difference was 2.85 U/L, with 95% limits of agreement from -0.56 to 6.25 U/L. Applying the same reference interval (11-34 U/L) to both methods yielded nine discordant classifications (6.6%), all upward with A-AST. Six of 114 specimens (5.3%) within the interval with AST2 exceeded the upper limit with A-AST. In an exploratory analysis, six of 11 specimens (54.5%) with AST2 results of 30-34 U/L exceeded the upper reference limit when measured with A-AST. The proportion above the limit increased from 10.2% to 14.6% (exact McNemar test, p = 0.031). This P5P matrix paradox shows that quality-assessment materials may not predict the direction of patient-result changes. The observed classification changes in paired specimens highlight the need to reassess reference interval applicability and alert clinicians to method-related shifts that could otherwise be interpreted as changes in clinical condition.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers