A Portable Electrical Response Verification System for Blood Glucose Meters Based on Dynamic Microcurrent Reproduction.
Wenshu Yue, Yifan Li, Jinhui Cai
College of Metrology Measurement and Instrument, China Jiliang University, Hangzhou 310018, China.
Key points
From the paper's conclusion- These results support the feasibility of controlled electrical response verification and short-term repeatability assessment of strip-based blood glucose meters.
Abstract
Portable blood glucose meters are commonly evaluated using test strips and control solutions, which assess the combined meter-strip-reagent response but are not designed to provide a directly controlled, pre-characterized electrical stimulus for repeated meter-side verification. This study developed a portable electrical response verification system based on dynamic microcurrent reproduction using a calibrated programmable current source. At each of two control solution levels, five 1000-point test strip current responses were acquired. Each response was reduced to 100 points by averaging every 10 consecutive samples, and the five reduced responses were then averaged point by point to obtain one level-specific mean reference waveform. Each mean waveform was converted into a 100-point DAC code table for STM32-controlled reproduction. The current source provided a measured output range of 0.016-9.829 μA. Five repeated reproductions per level yielded mean Pearson correlation coefficients of 0.9993 and 0.9995 and peak current CV values of 2.52% and 1.59%, respectively. The corresponding GA-3 meter (Sinocare Inc., Changsha, Hunan, China) readings differed from the mean original readings by 1.70% and 1.18%, with CV values of 1.81% and 1.29%. These results support the feasibility of controlled electrical response verification and short-term repeatability assessment of strip-based blood glucose meters.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Controlled In Vitro Characterization of the Dynamic Response of Continuous Glucose Monitoring Systems: Adaptation of a Programmable Flow Platform and Decomposition of Dynamic ErrorbioRxiv · 2026
- In Vitro Dynamic Characterization of a Continuous Glucose Monitoring System Using an Independently Measured Glucose Profile: Amplitude Transfer, Rate Transfer, and Consecutive-Day Agreement.Sensors (Basel, Switzerland) · 2026
- Single-use microneedle biosensor for detecting clinically relevant glucose in interstitial fluidBiosensors and Bioelectronics · 2026
- Análise comparativa de dois reagentes diferentes de glicose de marcas Mindray e Biosystem, mensurados no analisador bioquímico Mindray BS – 120Revista fisio&terapia. · 2026
- A-378 Performance Evaluation of the Roche Cobas® Pulse Point-of-Care GlucometerClinical Chemistry · 2026
- A self-calibrating microelectrode chip for glucose sensing in whole sturgeon blood.The Analyst · 2026