In Vitro Dynamic Characterization of a Continuous Glucose Monitoring System Using an Independently Measured Glucose Profile: Amplitude Transfer, Rate Transfer, and Consecutive-Day Agreement.
Elena V Khoroshun, Vladimir A Kozlov, Igor V Ivanov, Kuvat T Momynaliev
Federal State Organization "Russian Scientific and Research Institute for Medical Engineering", Federal Service for Surveillance in Healthcare (Roszdravnadzor), Kashirskoye Shosse 24, Bld. 16, 115478 Moscow, Russia.
Key points
From the paper's conclusion- We investigated the dynamic response of one factory-calibrated CGM system on a programmable in vitro flow platform, with glucose measured independently by a SUPER GL2 analyzer as comparator.
Abstract
Continuous glucose monitoring (CGM) sensors are evaluated almost exclusively in clinical studies. We investigated the dynamic response of one factory-calibrated CGM system on a programmable in vitro flow platform, with glucose measured independently by a SUPER GL2 analyzer as comparator. Twelve sensors were studied: four during verification of the measurement procedure (Stage 1) and eight tested on two consecutive days (Stage 2) under a 90 → 270 → 90 mg/dL profile. Amplitude transfer (KA) and apparent rising- and falling-phase slope transfer (Kup, Kdown) were calculated without lag correction. Stage 1 results were exploratory owing to an unexplained CGM displacement of approximately 35 min. The delivered profile differed from the programmed one in every Stage 2 run (upper plateaus 254-268 vs. 270 mg/dL). Mean KA was 0.731 and 0.731, Kup 0.749 and 0.746, and Kdown 0.729 and 0.719 on the two days; between sensors, KA ranged from 0.61-0.84. Near-zero mean paired changes reflected opposite group shifts (KA +0.026 and -0.025). Descriptive ICC(A,1) values of 0.919-0.952 (seven or eight pairs from two shared runs) indicate preserved between-sensor differences rather than established reproducibility. Bias (≈-50 mg/dL) and MARD (31-32%) apply only to these in vitro conditions; the findings do not establish clinical performance or long-term stability.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Controlled In Vitro Characterization of the Dynamic Response of Continuous Glucose Monitoring Systems: Adaptation of a Programmable Flow Platform and Decomposition of Dynamic ErrorbioRxiv · 2026
- International clinical opinion on transparency, standardisation, and calibration alignment in the performance evaluation of systems for continuous glucose monitoringDiabetes Obesity and Metabolism · 2026
- Toward standardized continuous glucose monitoring validation: lessons from the Sinocare iCan i3 accuracy study.Frontiers in clinical diabetes and healthcare · 2026
- Performance of the i-SENS CareSens Air CGM System in Compliance with the European CGM (eCGM) Clinical Performance Criteria: A Pooled Analysis.Diabetes therapy : research, treatment and education of diabetes and related disorders · 2026
- Impact of Missing Data and Monitoring Duration on Downstream Analyses in Continuous Glucose MonitoringDiabetes Care · 2026
- Glycemic Concordance of Simplera Sync ™ and Instinct ™ Sensors in MiniMed ™ 780G System Users: A Randomized Crossover Feasibility StudyDiabetes Technology & Therapeutics · 2026