Evaluation of the Bosphore HBV Quantification Kit: Categorical and Quantitative Agreement with the Aptima HBV Quant Assay for HBV DNA Detection and Quantification.
Massimiliano Guerra, Laura Dionisi, Claudia Colosimo, Alessandra Mistral De Pascali, Martina Brandolini, Ludovica Ingletto, Maria Sofia Montanari, Laura Grumiro,
Unit of Microbiology, The Greater Romagna Area Hub Laboratory, 47522 Cesena, Italy. · Department of Medical and Surgical Sciences (DIMEC), University of Bologna, 40138 Bologna, Italy.
Key points
From the paper's conclusion- Overall, 109/110 paired numerical results (99.1%; 95% CI, 95.04–99.98%) were within ±0.5 log10 IU/mL.
Abstract
Accurate hepatitis B virus (HBV) DNA quantification is essential for assessing viral replication and monitoring antiviral therapy. This retrospective method-comparison study evaluated the Bosphore HBV Quantification Kit against the Aptima HBV Quant Assay using 130 residual anonymized EDTA-plasma specimens, comprising 110 comparator-positive and 20 comparator-negative samples. Categorical agreement was assessed using positive, negative, and overall percent agreement (PPA, NPA, and OPA), while quantitative agreement was evaluated using a predefined ±0.5 log10 IU/mL criterion, Bland-Altman analysis, Passing-Bablok regression, and Lin's concordance correlation coefficient (CCC). PPA was 100% (110/110; 95% confidence interval [CI], 96.70-100.00%), NPA was 95.0% (19/20; 95% CI, 75.13-99.87%), and OPA was 99.23% (129/130; 95% CI, 95.79-99.98%). One confirmed qualitative discordance was observed. Overall, 109/110 paired numerical results (99.1%; 95% CI, 95.04-99.98%) were within ±0.5 log10 IU/mL. Bland-Altman analysis showed a mean bias of +0.124 log10 IU/mL, Passing-Bablok regression yielded Y = 0.107 + 1.019X, and Lin's CCC was 0.991. The Bosphore assay demonstrated high categorical and quantitative agreement with the comparator across the HBV DNA concentration range evaluated, supporting its potential suitability for quantitative HBV DNA testing within the analytical workflow assessed in this study.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Performance Evaluation of HBV RNA Real-Time PCR Kits for Quantitative Detection and Clinical Comparison Across Different Detection SystemsMicroorganisms · 2026
- A comparative viral load evaluation for human immunodeficiency virus, hepatitis B virus and hepatitis C virus of the Roche cobas 5800, Abbott Alinity m and Hologic Panther.Vox sanguinis · 2026
- Development and application of a real-time VPCR method for ultra-fast quantitative analysis and Visual Detection of HBV-DNA.Journal of clinical microbiology · 2026
- B-163 Clinical Evaluation of the Newly Approved Access HBsAg assay for Enhanced Detection, Workflow Efficiency, and Analytical Integrity in the United StatesClinical Chemistry · 2026
- Pilot Evaluation of Simplified HBV DNA Extraction-LAMP Workflows for Hepatitis B Molecular Testing in Resource-Limited Settings.Diagnostics (Basel, Switzerland) · 2026
- Analytical Agreement of the GeneXpert HIV-1 Viral Load Assay Compared with the Abbott Alinity m HIV-1 Assay.Diagnostics (Basel, Switzerland) · 2026