Analytical Agreement of the GeneXpert HIV-1 Viral Load Assay Compared with the Abbott Alinity m HIV-1 Assay.
Khulud A Alhazmi, Sulaiman Bani AbdelRahman, Hala Altarawneh, Karem Ibrahem, Waiel S Halabi, Abdulaziz Alsaedi, Turki Asiri, Bandar Hasan Saleh,
Department of Microbiology and Parasitology, Faculty of Medicine, Umm Al-Qura University, Makkah 24382, Saudi Arabia. · Department of Microbiology and Pathology, Faculty of Medicine, Mutah University, P.O. Box 7, Mutah, Al-Karak 61710, Jordan.
Key points
From the paper's conclusion- This study aimed to evaluate the qualitative and quantitative agreement between the GeneXpert HIV-1 Viral Load assay and the Abbott Alinity m HIV-1 assay.
Abstract
Background/Objectives: Early and accurate quantification of Human Immunodeficiency Virus type 1 (HIV-1) viral load is essential for monitoring antiretroviral therapy and guiding clinical management. Although conventional real-time polymerase chain reaction (PCR) platforms provide reliable viral load measurement, they often require centralized laboratories and longer turnaround times. Cartridge-based systems such as the GeneXpert HIV-1 Viral Load assay offer a simplified testing workflow and may facilitate decentralized molecular testing. This study aimed to evaluate the qualitative and quantitative agreement between the GeneXpert HIV-1 Viral Load assay and the Abbott Alinity m HIV-1 assay. Methods: A total of 102 residual clinical Ethylenediaminetetraacetic acid (EDTA) plasma specimens were analyzed using both the GeneXpert HIV-1 Viral Load assay and the Abbott Alinity m HIV-1 assay. Qualitative agreement was evaluated using positive percent agreement (PPA), negative percent agreement (NPA), overall agreement, and Cohen's κ coefficient. Quantitative agreement among specimens with paired quantifiable viral load results was assessed using Pearson's correlation and Bland-Altman analyses. Results: Overall qualitative agreement between the two assays was 95.1%, with a PPA of 92.6%, an NPA of 100.0%, and a Cohen's κ coefficient of 0.894, indicating almost perfect agreement. Among the 57 specimens with paired quantifiable viral load results, a very strong positive correlation was observed (Pearson's r = 0.9972; 95% CI: 0.9952-0.9984; p < 0.001). Bland-Altman analysis demonstrated a minimal mean bias of -0.018 log10 IU/mL, with 95% limits of agreement ranging from -0.175 to 0.139 log10 International Units per milliliter (IU/mL). All paired quantitative differences were within the clinically accepted agreement threshold of ±0.5 log10 IU/mL. Conclusions: The GeneXpert HIV-1 Viral Load assay demonstrated excellent qualitative and quantitative agreement with the Abbott Alinity m HIV-1 assay. These findings support the potential use of the GeneXpert HIV-1 Viral Load assay as an alternative testing platform for routine HIV-1 viral load monitoring in appropriate clinical settings.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Validation and performance of a rapid HIV-1 viral load assay.Microbiology spectrum · 2026
- A comparative viral load evaluation for human immunodeficiency virus, hepatitis B virus and hepatitis C virus of the Roche cobas 5800, Abbott Alinity m and Hologic Panther.Vox sanguinis · 2026
- Relative Diagnostic Yield of Tests for Disseminated Tuberculosis Among Inpatients With Human Immunodeficiency Virus.Open forum infectious diseases · 2026
- Large-scale evaluation of HIV-1 DNA drug resistance testing as a robust tool for clinical decision-making: A nationwide study in ChinaJournal of Pharmaceutical Analysis · 2025
- Evaluation of the diagnostic validity of the HIV VISITECT CD4 point-of-care rapid diagnostic test in Uganda.BMC infectious diseases · 2026
- Evaluation of the Bosphore HBV Quantification Kit: Categorical and Quantitative Agreement with the Aptima HBV Quant Assay for HBV DNA Detection and Quantification.Microorganisms · 2026