Analytical Agreement of the GeneXpert HIV-1 Viral Load Assay Compared with the Abbott Alinity m HIV-1 Assay.

Khulud A Alhazmi, Sulaiman Bani AbdelRahman, Hala Altarawneh, Karem Ibrahem, Waiel S Halabi, Abdulaziz Alsaedi, Turki Asiri, Bandar Hasan Saleh,

Department of Microbiology and Parasitology, Faculty of Medicine, Umm Al-Qura University, Makkah 24382, Saudi Arabia. · Department of Microbiology and Pathology, Faculty of Medicine, Mutah University, P.O. Box 7, Mutah, Al-Karak 61710, Jordan.

Key points

From the paper's conclusion
  • This study aimed to evaluate the qualitative and quantitative agreement between the GeneXpert HIV-1 Viral Load assay and the Abbott Alinity m HIV-1 assay.

Abstract

Background/Objectives: Early and accurate quantification of Human Immunodeficiency Virus type 1 (HIV-1) viral load is essential for monitoring antiretroviral therapy and guiding clinical management. Although conventional real-time polymerase chain reaction (PCR) platforms provide reliable viral load measurement, they often require centralized laboratories and longer turnaround times. Cartridge-based systems such as the GeneXpert HIV-1 Viral Load assay offer a simplified testing workflow and may facilitate decentralized molecular testing. This study aimed to evaluate the qualitative and quantitative agreement between the GeneXpert HIV-1 Viral Load assay and the Abbott Alinity m HIV-1 assay. Methods: A total of 102 residual clinical Ethylenediaminetetraacetic acid (EDTA) plasma specimens were analyzed using both the GeneXpert HIV-1 Viral Load assay and the Abbott Alinity m HIV-1 assay. Qualitative agreement was evaluated using positive percent agreement (PPA), negative percent agreement (NPA), overall agreement, and Cohen's κ coefficient. Quantitative agreement among specimens with paired quantifiable viral load results was assessed using Pearson's correlation and Bland-Altman analyses. Results: Overall qualitative agreement between the two assays was 95.1%, with a PPA of 92.6%, an NPA of 100.0%, and a Cohen's κ coefficient of 0.894, indicating almost perfect agreement. Among the 57 specimens with paired quantifiable viral load results, a very strong positive correlation was observed (Pearson's r = 0.9972; 95% CI: 0.9952-0.9984; p < 0.001). Bland-Altman analysis demonstrated a minimal mean bias of -0.018 log10 IU/mL, with 95% limits of agreement ranging from -0.175 to 0.139 log10 International Units per milliliter (IU/mL). All paired quantitative differences were within the clinically accepted agreement threshold of ±0.5 log10 IU/mL. Conclusions: The GeneXpert HIV-1 Viral Load assay demonstrated excellent qualitative and quantitative agreement with the Abbott Alinity m HIV-1 assay. These findings support the potential use of the GeneXpert HIV-1 Viral Load assay as an alternative testing platform for routine HIV-1 viral load monitoring in appropriate clinical settings.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Analytical Agreement of the GeneXpert HIV-1 Viral Load Assay Compared with the Abbott Alinity m HIV-1 Assay — Science Skim