B-163 Clinical Evaluation of the Newly Approved Access HBsAg assay for Enhanced Detection, Workflow Efficiency, and Analytical Integrity in the United States

Jennifer Kalamatas, Saeed Jortani, Robert Christenson, Fred Apple, Marc Turini, Danielle Maher

Key points

From the paper's conclusion
  • Unlike the comparator assay, which experienced instances of carryover during clinical evaluation, the Access HBsAg assay, utilizing the DxI 9000's disposable tip technology, exhibited no vulnerability to carryover, ensuring the integrity and reliability of every result.

Abstract

Background Accurate and timely detection of Hepatitis B surface antigen (HBsAg) is crucial for the diagnosis, screening, and management of Hepatitis B virus (HBV) infection. Traditional HBsAg testing workflows often involve multiple steps, including repeat and confirmatory testing, which can delay results and increase laboratory burden and costs. There is a continuous need for assays that offer improved sensitivity, efficiency, and robustness. Methods This clinical study evaluated the performance of the newly FDA approved Beckman Coulter Access HBsAg assay on the DxI 9000 Immunoassay Analyzer. It involved testing 3,614 samples from adult, pregnant, and pediatric patients from at-risk populations with or without signs and symptoms with various stages of HBV infection at three US clinical trial sites. The assay’s performance was directly compared against an FDA-approved HBsAg comparator assay. The study focused on assessing clinical performance, the utility of a novel high positive algorithm for result interpretation, and the assay's susceptibility to analytical carryover. Results The Access HBsAg assay demonstrated high clinical performance with an overall PPA of 99.5% (n=417; 95% CI: 98.3-99.9) and NPA of 99.7% (n=3197; 95% CI: 99.5-99.9). Stage-specific performance was also high across different HBV disease stages, with a PPA of 100.0% (n=3; 95%CI: 43.9-100.0) for early acute infection, 100.0% (n=17; 95% CI: 81.6-100.0) for acute infection, and 99.7% (n=394; 95%CI: 98.6-100.0) for chronic infection. The assay exhibited improved sensitivity compared with the FDA-approved comparator, successfully detecting HBsAg in 3 samples that were not detected by the comparator. A novel high positive algorithm proved a key advancement: samples yielding an initial HBsAg result >100 S/CO were consistently confirmed positive. During this trial, 94.3% (399 of 423) of initially reactive samples fell into the = 100.00 S/CO range and 100.0% (399/399) were confirmed positive using the Access HBsAg Confirmatory assay, thereby eliminating the need for repeat or confirmatory testing for these high-titer samples. This significantly streamlines laboratory workflow and accelerates diagnostic turnaround time. Unlike the comparator assay, which experienced instances of carryover during clinical evaluation, the Access HBsAg assay, utilizing the DxI 9000's disposable tip technology, exhibited no vulnerability to carryover, ensuring the integrity and reliability of every result. Conclusion The Beckman Coulter Access HBsAg assay offers significant advancements for HBsAg detection. Its enhanced sensitivity provides a more comprehensive detection capability, identifying samples that might otherwise be missed. The novel high positive algorithm optimizes laboratory efficiency by reducing the need for costly and time-consuming repeat or confirmatory testing. Coupled with its inherent resistance to carryover due to disposable tips on the DxI 9000, this assay delivers a highly reliable, efficient, and sensitive solution for HBsAg detection, ultimately benefiting patient care and laboratory operations.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

B-163 Clinical Evaluation of the Newly Approved Access HBsAg assay for Enhanced Detection, Workflow Efficiency, and Analytical Integrity in the United States — Science Skim