Performance Evaluation of HBV RNA Real-Time PCR Kits for Quantitative Detection and Clinical Comparison Across Different Detection Systems
Xin Ai, Xin Wang, Runling Zhang, Feng Liu, Hongsong Chen, Yuyuan Jia, Haiying Zhang, Huiying Rao
Key points
From the paper's conclusion- Kit A met the predefined analytical performance requirements under this laboratory’s conditions, but quantitative bias and low-concentration detection differences between the two HBV RNA testing systems underscore the need to standardize calibrators, assay target regions, sample pretreatment, and amplification platforms to improve result comparability.
Abstract
Serum hepatitis B virus (HBV) RNA has emerged as a potential biomarker of residual viral transcriptional activity in patients with chronic hepatitis B (CHB). However, insufficient assay standardization and variability across testing systems limit the reliability and comparability of HBV RNA measurements. This study aimed to verify the analytical performance of a commercial HBV RNA detection assay (Kit A) and assess its methodological comparability with another commercial assay (Kit B) using clinical serum samples under different testing conditions. This study comprised analytical performance verification of Kit A, clinical sample comparison between Kit A and Kit B using contemporaneous fresh samples, long-term frozen samples, and non-contemporaneous refrigerated samples, and an exploratory comparison of amplification platforms. Both kits were tested in parallel on the same amplification instrument using different nucleic acid extraction platforms, with operators blinded to the other reagent’s results, and analyses included qualitative agreement, quantitative bias, and low-concentration detection concordance. Kit A met all predefined analytical performance criteria, including trueness (mean biases of −0.06 and −0.03 log10 copies/mL), linearity (R2 = 0.9997; slope 1.008), the 50 copies/mL limit of detection (20/20 detected), precision (CVs of 1.22% and 2.29%), interference resistance (|B| < 7.5%), and analytical specificity (no HBV RNA target detected in 30 tests). In clinical comparisons, qualitative agreement was 88.7% for contemporaneous fresh samples, 92.5% for long-term frozen samples, and 72.9% for non-contemporaneous refrigerated samples, with discordant results concentrated in low-concentration samples; among samples with Kit B results ≥ 300 copies/mL, Kit A showed strong quantitative correlation but positive systematic bias (mean biases of 0.428 and 0.668 log10 copies/mL, respectively, with proportional bias on Passing–Bablok regression). Exploratory platform comparison suggested a higher detection rate on the Applied Biosystems 7500 Fast Real-Time PCR System and the Applied Biosystems™ QuantStudio™ 5 Real-Time PCR System, but the limited data were insufficient to determine overall platform differences. Kit A met the predefined analytical performance requirements under this laboratory’s conditions, but quantitative bias and low-concentration detection differences between the two HBV RNA testing systems underscore the need to standardize calibrators, assay target regions, sample pretreatment, and amplification platforms to improve result comparability.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Evaluation of the Bosphore HBV Quantification Kit: Categorical and Quantitative Agreement with the Aptima HBV Quant Assay for HBV DNA Detection and Quantification.Microorganisms · 2026
- Development and application of a real-time VPCR method for ultra-fast quantitative analysis and Visual Detection of HBV-DNA.Journal of clinical microbiology · 2026
- Integrated recombinase-aided amplification–invader system: a closed-tube visual platform for hepatitis B virus detectionFrontiers in Cellular and Infection Microbiology · 2026
- A novel highly specific qPCR assay for distinguishing HBV cccDNA from rcDNA following bisulfite pretreatment.Journal of virology · 2026
- B-163 Clinical Evaluation of the Newly Approved Access HBsAg assay for Enhanced Detection, Workflow Efficiency, and Analytical Integrity in the United StatesClinical Chemistry · 2026
- Multicenter evaluation of an innovative fully-automated molecular diagnostics system.Clinical chemistry and laboratory medicine · 2026