Influence of Corneal Hysteresis on Agreement Between Corneal-Compensated, Non-Contact, and Rebound Tonometry
Florina Vultur, Elena Andreea Manescu, Karin Ursula Horvath
Key points
From the paper's conclusion- These findings suggest that considering CH in addition to CCT may support a more accurate interpretation of IOP measurements in biomechanically vulnerable eyes.
Abstract
Background/Objectives: Corneal hysteresis (CH) may provide information beyond central corneal thickness (CCT) when examining differences between tonometric measurements. This study assessed associations between CH and inter-tonometer differences after adjustment for CCT and age. Methods: This cross-sectional study included 134 eyes from 67 patients attending follow-up one month after cataract surgery in one eye. Both eyes were examined using non-contact tonometry (NCT), iCare Pro, and the Ocular Response Analyzer (ORA). CCT was measured using the Topcon CT-1P. Agreement was assessed using Bland–Altman analysis, including proportional bias and regression-based limits of agreement. Multivariable models and comparisons across three CH groups used patient-clustered robust standard errors. Complete observations were used separately for each analysis. Results: Mean biases for IOPcc–NCT, IOPcc–iCare, and NCT–iCare were 0.15 mmHg (95% LoA, −5.15, 5.44), 1.68 mmHg (−5.41, 8.77), and 1.18 mmHg (−4.42, 6.78), respectively. Significant proportional bias was observed only for IOPcc–NCT. After adjustment for CCT and age, each 1-mmHg decrease in CH was associated with increases of 1.56, 2.23, and 0.39 mmHg in the respective differences (p < 0.001, p < 0.001, and p = 0.042). In the low CH group, IOPcc exceeded NCT and iCare by averages of 2.50 and 5.51 mmHg, respectively. Conclusions: Lower CH was associated with more positive inter-tonometer differences, particularly those involving IOPcc. Wide limits of agreement do not support assuming interchangeability. While mean differences were small in the overall cohort, they became clinically relevant in eyes with low CH. These findings suggest that considering CH in addition to CCT may support a more accurate interpretation of IOP measurements in biomechanically vulnerable eyes.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Precision and reliability in tonometry: Comparison of Corvis ST and iCare in glaucoma patients on the Ecuadorian coast.Journal francais d'ophtalmologie · 2026
- Dynamic corneal responses from successive tonometry at different resting intervals using the Corvis ST in young myopesFrontiers in Bioengineering and Biotechnology · 2026
- Comparison of Horizontal and Vertical Tonovet Plus Clinical Measurement Configurations for Intraocular Pressure Assessment in Conscious CatsVeterinary Ophthalmology · 2026
- Differences in intraocular pressure measurements by tonometer type in children with glaucoma and suspected glaucoma.Journal of AAPOS : the official publication of the American Association for Pediatric Ophthalmology and Strabismus · 2026
- Role of Corneal Biomechanics in the Pathogenesis, Diagnosis, and Management of Glaucoma.Journal of ophthalmology · 2026
- Transient Corneal Response After Goldmann Three-Mirror Gonioscopy and Its Limited Effect on Optical Biometry and Calculated Intraocular Lens Power: A Prospective Within-Subject Controlled Observational Study.Medicina (Kaunas, Lithuania) · 2026