Comparison of Horizontal and Vertical Tonovet Plus Clinical Measurement Configurations for Intraocular Pressure Assessment in Conscious Cats
Mehmet Nur Çetin, Yusuf Sinan Şirin, Doğukan Polat, Batuhan Neyse, Latif Emrah Yanmaz
Key points
From the paper's conclusion- These findings indicate that horizontal and vertical Tonovet Plus clinical measurement configurations may yield different IOP values in conscious cats.
Abstract
Rebound tonometry is widely used for measuring intraocular pressure (IOP) in cats; however, differences in clinical tonometer measurement configuration may contribute to variability in recorded values. This prospective repeated‐measures study compared IOP measurements obtained using horizontal and vertical Tonovet Plus clinical measurement configurations in 32 client‐owned, systemically and ophthalmologically normal conscious cats. Intraocular pressure was measured in the right eye while the probe was directed toward the center of the cornea and positioned 4–8 mm from the corneal surface. The order of horizontal and vertical configurations was randomized for each cat. Measurements were obtained by one experienced veterinary ophthalmologist and a trained student operator group. In the primary linear mixed‐effects analysis accounting for repeated measurements within cats, horizontal configuration was associated with an estimated 2.41‐mmHg higher IOP than vertical configuration ( β = 2.41 mmHg; 95% CI, 1.39–3.44; p < 0.001). The configuration‐associated difference was also observed separately for the experienced instructor (mean difference, 2.81 mmHg; 95% CI, 1.91–3.71; p < 0.001) and across the trained student operator group (2.00 mmHg; 95% CI, 0.81–3.19; p = 0.002). These findings indicate that horizontal and vertical Tonovet Plus clinical measurement configurations may yield different IOP values in conscious cats. As configuration changes may affect instrument handling, restraint, head positioning, and animal responses, the difference cannot be attributed solely to tonometer orientation or mechanics. Consistent configuration is therefore recommended when comparing serial IOP measurements in cats. Further controlled studies are needed to distinguish mechanical, ergonomic, and other examination‐related mechanisms.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Multi-angle validation of the iFalcon V700 rebound tonometer against implantable pressure sensor in ex-vivo rabbit eyesScientific Reports · 2026
- Changing tonometers, changing intraocular pressure? Rethinking intraocular pressure interpretation in dogs.The Journal of small animal practice · 2026
- Ex Vivo Validation of the TONOVET Plus and TONOVET Pro in Cynomolgus Monkeys (Macaca fascicularis): Comparison of Canine and Lapine Calibration Modes.Veterinary ophthalmology · 2026
- Precision and reliability in tonometry: Comparison of Corvis ST and iCare in glaucoma patients on the Ecuadorian coast.Journal francais d'ophtalmologie · 2026
- A Method for Measuring Intraocular Pressure in Wild-Type Zebrafish Using Rebound Tonometry: Characterization, Sex Differences, and Long-Term Repeatability.Zebrafish · 2026
- Inter‐Device Consistency Between iFalcon and iCare in Mechanically Pressurized Ex Vivo Rat EyesVeterinary Ophthalmology · 2026