Differences in intraocular pressure measurements by tonometer type in children with glaucoma and suspected glaucoma.
Yeabsira Mesfin, Keith D Miller, Ying Han, Julius T Oatts
Department of Ophthalmology, University of California, San Francisco School of Medicine, San Francisco, California. · Division of Ophthalmology, Children's Hospital of Philadelphia, Scheie Eye Institute, Perelman School of Medicine at the University of Pennsylvania, Philadelphia, Pennsylvania. Electronic address: oattsj2@chop.edu.
Key points
From the paper's conclusion- In our study cohort of children with glaucoma and suspected glaucoma, IOP measurements obtained with pneumotonometry and rebound tonometry measured higher than GAT but showed good correlation.
Abstract
PURPOSE: To compare methods for obtaining intraocular pressure (IOP) measurements in children with glaucoma or suspected glaucoma. METHODS: The medical records of children (<18 years) with glaucoma or suspected glaucoma with IOP measured by Goldmann applanation tonometry (GAT) and either pneumotonometry or rebound tonometry at a single institution from March 2018 to August 2024 were reviewed retrospectively. The difference between measurements was recorded for each eye. Correlation and Bland-Altman analysis were used to evaluate the level of agreement between tonometers. A linear mixed effects model identified characteristics associated with IOP differences. RESULTS: A total of 157 eyes of 89 patients were included. GAT was used in all eyes, pneumotonometry in 79 eyes, and rebound tonometry in 78 eyes. The mean IOP by GAT, pneumotonometry, and rebound tonometry was 17.3 ± 5.5 mm Hg, 18.8 ± 5.5 mm Hg, and 20.4 ± 6.0 mm Hg, respectively. Pneumotonometry and rebound tonometry measured higher than GAT with a mean difference of 1.4 ± 3.2 mm Hg (limits of agreement [LOA]: 7.6 mm Hg, -4.3 mm Hg) and 3.1 ± 3.5 mm Hg (LOA: 10.0 mm Hg, -3.9 mm Hg). Correlations between GAT and both pneumotonometry (R = 0.83, P < 0.0001) and rebound tonometry (R = 0.91, P < 0.0001) were strong. Pneumotonometry was associated with a smaller difference from GAT measurements than rebound tonometry (estimate, -1.54 mm Hg). The difference between measurements from rebound tonometry and GAT was influenced by CCT (estimate, 2.35 mm Hg greater discrepancy per 100 μm [P = 0.004]). CONCLUSIONS: In our study cohort of children with glaucoma and suspected glaucoma, IOP measurements obtained with pneumotonometry and rebound tonometry measured higher than GAT but showed good correlation.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Comparison of Intraocular Pressure Measurements Using iCare Rebound Tonometry and Goldmann Applanation Tonometry: A Systematic Review and Meta-Analysis.Journal of glaucoma · 2026
- Precision and reliability in tonometry: Comparison of Corvis ST and iCare in glaucoma patients on the Ecuadorian coast.Journal francais d'ophtalmologie · 2026
- Agreement between iCare HOME and Goldmann applanation tonometry: systematic review and meta-analysis.Canadian journal of ophthalmology. Journal canadien d'ophtalmologie · 2026
- Influence of Corneal Hysteresis on Agreement Between Corneal-Compensated, Non-Contact, and Rebound TonometryJournal of Clinical Medicine · 2026
- Comparison of Horizontal and Vertical Tonovet Plus Clinical Measurement Configurations for Intraocular Pressure Assessment in Conscious CatsVeterinary Ophthalmology · 2026
- Clinical evaluation of corneal elasticity and true IOP using a dual-applanation tonometerFrontiers in Ophthalmology · 2026