Precision and reliability in tonometry: Comparison of Corvis ST and iCare in glaucoma patients on the Ecuadorian coast.
R E Alcibar-Viteri, V D Moreira-Pico, K Montes-Escobar, K E Intriago-Flores, M C Pérez Pirela, J P Duran-Ospina, A Siteneski, N R Maddela
Instituto de la Visión, Portoviejo 130105, Ecuador. · Departamento de Formación y Desarrollo Científico en Ingeniería, Facultad de Ingeniería, Ciencia y Tecnología, Universidad Bernardo O'Higgins, Santiago, Chile; Departamento de Ciencias Agrícolas, Facultad de Ingeniería Agrícola, Universidad Técnica de Manabí, Portoviejo 130150, Ecuador.
Key points
From the paper's conclusion- The biomechanically corrected bIOP yielded the most consistent, corneal-thickness-independent estimates of IOP. Although iCare remains convenient for screening, bIOP measurements appear superior for precise monitoring and therapeutic decision-making in glaucoma care.
Abstract
BACKGROUND: This study aimed to compare IOP or tonometric measurements obtained with the Corvis ST and the iCare rebound tonometer in glaucoma patients seen at an ophthalmology center on the Pacific coast of Ecuador. METHODS: This study retrospectively reviewed clinical records obtained between June 2023 and January 20 across six institutional sites. A total of 434 patients with bilateral glaucoma met the inclusion criteria. Each participating site had a single trained examiner responsible for all measurements during the study period, ensuring consistency within each location. Mean differences between instruments were evaluated using repeated measures analysis, and inter-method agreement was assessed using Bland-Altman plots. RESULTS: The biomechanically corrected IOP (bIOP) from the Corvis ST was lower than the iCare IOP measurements in 79% of eyes; the overall mean difference was 1.7 millimeters of mercury (standard deviation 2.8; P<0.001). The limits of agreement for the corrected mode ranged from -6.1 to +2.7 millimeters, compared with -9.8 to +3.8 for the uncorrected setting. Variability, assessed by the coefficient of repeatability, was 32% lower with bIOP than with the rebound tonometer. Central corneal thickness correlated with uIOP readings (uncorrected intraocular pressure from the Corvis ST) (r=0.32, P<0.01), but not with the corrected values or with the iCare measurements. Right and left eyes showed consistent patterns. CONCLUSION: The biomechanically corrected bIOP yielded the most consistent, corneal-thickness-independent estimates of IOP. Although iCare remains convenient for screening, bIOP measurements appear superior for precise monitoring and therapeutic decision-making in glaucoma care.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Corneal Biomechanical Metrics by Corvis ST in a Mexican Population: Normative Data and Population-Specific Implications for Intraocular Pressure Assessment.Journal of refractive surgery (Thorofare, N.J. : 1995) · 2026
- Influence of Corneal Hysteresis on Agreement Between Corneal-Compensated, Non-Contact, and Rebound TonometryJournal of Clinical Medicine · 2026
- Comparison of Intraocular Pressure Measurements Using iCare Rebound Tonometry and Goldmann Applanation Tonometry: A Systematic Review and Meta-Analysis.Journal of glaucoma · 2026
- Differences in intraocular pressure measurements by tonometer type in children with glaucoma and suspected glaucoma.Journal of AAPOS : the official publication of the American Association for Pediatric Ophthalmology and Strabismus · 2026
- Inter-eye comparison of corneal stiffness parameters in primary open-angle glaucoma patients with unilateral damage.Graefe's archive for clinical and experimental ophthalmology = Albrecht von Graefes Archiv fur klinische und experimentelle Ophthalmologie · 2026
- Corneal Biomechanics in Glaucoma: Current Evidence for Diagnosis, Risk Stratification, and Prognosis.Clinical ophthalmology (Auckland, N.Z.) · 2026