Primary Stability of Implants Inserted Using Different Osseodensification Systems in Low-Density Bone: An In Vitro Study
André Luís Onodera, Alexandre Wanderley Alécio, Gustavo Batista Grolli Klein, Sheila Cortelli, Rogério de Lima Romeiro
Faculdade São Leopoldo Mandic · Universidade de Taubaté
Key points
From the paper's conclusion- Weak-to-moderate correlations between variables in both analyses indicated that higher insertion torque did not necessarily translate into greater stability.
Abstract
Background/Objectives: It remains unclear which drilling strategy is most effective for maximizing mechanical stability in low-density bone and whether high insertion torque is determinative. The aim of this in vitro study was to compare the primary stability of implants placed using different drilling protocols—conventional (CV), undersized (US), and osseodensification (OD). Three osseodensification systems—Versah burs (V), Bone Reamer Drills (WF), and Master Conical Densifiers (DSP)—were also compared. Methods: A set of 11 blocks was used for the drilling protocol comparison (CV, US, OD) and a separate set of 11 blocks was used for the osseodensification system comparison (V, WF, DSP). External-hexagon implants were inserted epicrestally. Insertion torque was measured using a torque meter, and implant stability quotients (ISQs) were assessed through resonance frequency analysis. Results: ISQ for OD was significantly higher than that for CV but statistically similar to that for US, whereas insertion torque for OD was significantly higher than that for both US and CV. A weak correlation was found between variables for CV and US, and a moderate one was observed for OD. Both WF and DSP showed significantly higher ISQ values than V. Insertion torque for DSP was significantly higher than that for both WF and V. A moderate correlation was found between variables for DSP and V, and a weak one for WF. Conclusions: In this invitro study, the OD protocol performed better than CV in terms of ISQ and better than both CV and US in terms of insertion torque. WF and DSP outperformed V in ISQ, whereas DSP yielded the highest insertion torque. Weak-to-moderate correlations between variables in both analyses indicated that higher insertion torque did not necessarily translate into greater stability.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Effect of Undersized Drilling and Sequential Osseodensification on Primary Implant Stability in Simulated Bone Types III and IV: An In Vitro Study.International journal of dentistry · 2026
- Influence of Different Drilling Protocols for Primary Stability at Low-Density Bone Detected by Bone Density Measurement Drill.Clinical and experimental dental research · 2026
- Effects of osteotomy preparation protocols and implant macrogeometry on primary stability in an ex vivo bovine-rib modelBMC Oral Health · 2026
- Evaluation of the Effects of Dental Implant Dimension and Design on Primary Implant Stability Using Different Stability Measurement Devices.Journal of clinical medicine · 2026
- Effect of Progressive Versus Standard Implant Thread Designs on Primary Stability and Insertion Torque: An In Vitro StudyThe International Journal of Oral & Maxillofacial Implants · 2026
- Influence of Implant Macrodesign on Primary Stability Across Different Bone Density Models: A Comparative In Vitro Study.Journal of functional biomaterials · 2026