Influence of Different Drilling Protocols for Primary Stability at Low-Density Bone Detected by Bone Density Measurement Drill.
Yusuke Morimoto, Reiko Kobatake, Kazuya Doi, Yoshifumi Oki, Kaien Wakamatsu, Tomoko Izumikawa, Kazuhiro Tsuga
Department of Advanced Prosthodontics, Hiroshima University Graduate School of Biomedical and Health Sciences, Hiroshima, Japan.
Key points
From the paper's conclusion- DTV reflects the existing bone volume of low-density tissue and is correlated with primary stability. Intraoperative assessment of drilling torque would facilitate the selection of appropriate drilling protocols to achieve primary stability. Bone density evaluation by measuring the DTV may provide a clinically useful method for the intraoperative evaluation at low-density implant sites.
Abstract
OBJECTIVES: Intraoperative bone assessment is critical for achieving primary implant stability, particularly at low-density bone sites. Drilling torque value (DTV) has been proposed as an indicator of bone density and may help guide drilling preparation protocols to improve primary stability. However, the relationship between DTV and primary stability, especially histological bone parameters in biological bone, remains unclear. This study aimed to clarify the influence of different drilling protocols for primary stability at low-density bone detected by DTVs in biological cancellous bones. MATERIALS AND METHODS: Cancellous bone blocks obtained from bovine ribs were used. A specially designed torque measurement drill was used to assess bone density intraoperatively. Based on the DTVs, low-density bone sites corresponding to Misch D4 and D5 were selected. The dental Implants were placed using either a normal or undersized drilling protocol. The DTVs, insertion torque values (ITVs), and implant stability quotient (ISQ) were recorded. Histological evaluation was performed to assess the bone area fraction (BAF) around the implant. RESULTS: The ITVs were lower than the clinically recommended range for primary stability in both groups; however, they were significantly higher in the undersized group than in the normal-sized group. The DTVs positively correlated with the ITVs in both groups. A positive correlation between DTV and ISQ was observed only in the normal-sized group, whereas ITV was positively correlated with ISQ in both groups. Histological analysis revealed no significant difference in BAF between the two groups. BAF positively correlated with DTV, ITV, and ISQ. CONCLUSIONS: DTV reflects the existing bone volume of low-density tissue and is correlated with primary stability. Intraoperative assessment of drilling torque would facilitate the selection of appropriate drilling protocols to achieve primary stability. Bone density evaluation by measuring the DTV may provide a clinically useful method for the intraoperative evaluation at low-density implant sites.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Primary Stability of Implants Inserted Using Different Osseodensification Systems in Low-Density Bone: An In Vitro StudyDentistry Journal · 2026
- Effect of Undersized Drilling and Sequential Osseodensification on Primary Implant Stability in Simulated Bone Types III and IV: An In Vitro Study.International journal of dentistry · 2026
- Influence of Implant Macrodesign on Primary Stability Across Different Bone Density Models: A Comparative In Vitro Study.Journal of functional biomaterials · 2026
- Effects of osteotomy preparation protocols and implant macrogeometry on primary stability in an ex vivo bovine-rib modelBMC Oral Health · 2026
- Evaluation of the Effects of Dental Implant Dimension and Design on Primary Implant Stability Using Different Stability Measurement Devices.Journal of clinical medicine · 2026
- Influence of Implant Macrogeometry on Primary Stability in Different Polyurethane Bone-Density Models: An In Vitro Study.Bioengineering (Basel, Switzerland) · 2026