Influence of Implant Macrodesign on Primary Stability Across Different Bone Density Models: A Comparative In Vitro Study.
Lidija Veljkovic, Miljana Nedeljkovic, Jovana Milanovic, Aleksandra Arnaut, Momir Stevanovic, Gvozden Rosic, Dragica Selakovic, Nemanja Jovicic,
Department of Dentistry, Faculty of Medical Sciences, University of Kragujevac, 34000 Kragujevac, Serbia. · Department of Physiology, Faculty of Medical Sciences, University of Kragujevac, 34000 Kragujevac, Serbia.
Key points
From the paper's conclusion- Conclusions: Implant macrodesign influences primary stability, with AnyRidge demonstrating the highest ITV and ISQ values in D1 as well as D4.
Abstract
BACKGROUND: Implant macrogeometry and bone density are major determinants of primary stability. This study aimed to evaluate primary stability across different implant macrodesigns and block models. MATERIALS AND METHODS: A total of 153 implants were divided into Group 1 (AnyRidge), Group 2 (BlueDiamond RC), and Group 3 (BlueDiamond NC) (n = 51 each). Each implant was sequentially tested in synthetic D1-D4 bone models. Insertion torque values (ITV) and implant stability quotient (ISQ) were recorded. RESULTS: In D1, ITV was highest in Group 1 (69.51 ± 1.50 Ncm), followed by Group 2 (53.82 ± 6.75 Ncm) and Group 3 (49.31 ± 7.00 Ncm) (p < 0.01, p < 0.001). In D4, ITV decreased to 34.71 ± 3.23, 19.90 ± 3.08, and 11.18 ± 2.14 Ncm, respectively (p < 0.001). D1 ISQ values were 84.53 ± 2.51, 72.08 ± 3.69, and 66.33 ± 2.63, respectively (p < 0.001). In D4, ISQ values were 60.29 ± 3.84, 52.86 ± 3.73, and 52.18 ± 2.48, respectively (p < 0.001); Groups 2 and 3 did not differ significantly. CONCLUSIONS: Implant macrodesign influences primary stability, with AnyRidge demonstrating the highest ITV and ISQ values in D1 as well as D4. Further clinical studies would be beneficial to support these findings.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Influence of Implant Macrogeometry on Primary Stability in Different Polyurethane Bone-Density Models: An In Vitro Study.Bioengineering (Basel, Switzerland) · 2026
- Evaluation of the Effects of Dental Implant Dimension and Design on Primary Implant Stability Using Different Stability Measurement Devices.Journal of clinical medicine · 2026
- Primary Stability of Implants Inserted Using Different Osseodensification Systems in Low-Density Bone: An In Vitro StudyDentistry Journal · 2026
- Effect of Progressive Versus Standard Implant Thread Designs on Primary Stability and Insertion Torque: An In Vitro StudyThe International Journal of Oral & Maxillofacial Implants · 2026
- Effects of osteotomy preparation protocols and implant macrogeometry on primary stability in an ex vivo bovine-rib modelBMC Oral Health · 2026
- Influence of Reduced Cortical Bone Compression by Implant Macrogeometry on Peri-Implant Bone Healing: An In Vitro and In Vivo Experimental StudyJournal of Functional Biomaterials · 2026