An Autonomous Robotic Phlebotomy Device: Extended Analytical Validation and Real-World Hemolysis Assessment.
Jeroen A Roest, Luuk F P Giesen, Thijs C van Holten, Martine M L Deckers, Helena Chon, Frederique M A Koopman, Michael A Fouraux, Judith M E P Gillis,
Department of Clinical Affairs, Vitestro, Europalaan 500, 3526 KS Utrecht, The Netherlands. · Department of Clinical Chemistry, St. Antonius Hospital, Koekoekslaan 1, 3435 CM Nieuwegein, The Netherlands.
Key points
From the paper's conclusion- The results further substantiate that the ARPD could improve patient care by reducing preanalytical errors.
Abstract
Background/Objectives: Automated phlebotomy may improve preanalytical quality and access to venous blood testing. Prior work evaluating the Aletta Autonomous Robotic Phlebotomy Device (ARPD) demonstrated analytical equivalence. This study expanded the initial analytical validation to additional blood collection tubes and analytes and also assessed real-world hemolysis rates. Methods: The study was conducted in three hospital outpatient phlebotomy departments, including a site with a high prevalence of (hemato-)oncology patients. The within-subject analytical validation study included Greiner VACUETTE (Cohort 1) and BD Vacutainer (Cohort 2) serum, lithium heparin, and EDTA tubes. Equivalence of potassium, lactate dehydrogenase, hemolysis index, and complete blood count (CBC) with differential between paired ARPD-collected and manually collected samples was assessed using Passing-Bablok regression. Real-world ARPD hemolysis (>0.5 g/L free hemoglobin) rates were determined in a separate cohort (Cohort 3) and compared with site-specific manual hemolysis rates. Results: In the analytical validation study, 119 patients were included in the analysis (Cohort 1, n = 52; Cohort 2, n = 67). All tests met the predefined acceptability criteria. The hemolysis index was statistically significantly lower in ARPD-collected lithium heparin samples in Cohort 1. In real-world use (1368 ARPD venipunctures), the ARPD hemolysis rate was 0.0% (95% CI: 0.0-0.3%), significantly lower than manual phlebotomy (1.6%, p = 0.025). Conclusions: The study demonstrated analytical equivalence in a clinically complex population and found reduced ARPD hemolysis rates. The results further substantiate that the ARPD could improve patient care by reducing preanalytical errors.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- A-277 Improving Sample Integrity: Preclinical Evidence Demonstrating Reduced Hemolysis with Novel Blood Collection DeviceClinical Chemistry · 2026
- Phlebotomy technique and its impact on haemolysis in the emergency department.Annals of clinical biochemistry · 2026
- Clinical evaluation of an intelligent venipuncture robotic system: performance, patient acceptance, and pre-analytical outcomes.Clinical chemistry and laboratory medicine · 2026
- Preanalytical interventions to reduce hemolysis and improve specimen quality: a systematic review and meta-analysis.Laboratory medicine · 2026
- Capillary Blood Collection Methods: An assessment of different collections technologies compared to traditional venous sampling.Annals of clinical biochemistry · 2026
- A-292 Innovation in Home Phlebotomy Services Using an In-Vehicle CentrifugeClinical Chemistry · 2026