Clinical Chemistry

A-292 Innovation in Home Phlebotomy Services Using an In-Vehicle Centrifuge

Daiane Gallo, Alline Nobrega, Gabriel Neto, Mauro Pinto

Key points

From the paper's conclusion
  • The results of the second phase evaluated with potassium showed that 15% of the patients had values impacted by late centrifugation after the route, changing the clinical judgment.

Abstract

Background One of the main purposes of laboratory tests is to guide clinical reasoning. For this, it is necessary to ensure quality in all laboratory processes, especially in the pre-analytical phase, which originates up to 70% of laboratory errors. Home collection brought convenience to the patient, but also technical challenges, such as hemolysis control due to the delay in centrifugation. This study aims to evaluate the technical performance and care benefits of the adoption of a vehicular centrifuge for home collection. This study was conducted in a large outpatient clinical laboratory located in Curitiba, Paraná, Brazil, and aims to evaluate the technical performance and clinical benefits of implementing an in-vehicle centrifuge for home phlebotomy services. Methods The LW Scientific E8F vehicular centrifuge (fixed) was validated in two stages. In the first, three samples were collected in a tube with separating gel from 25 volunteers, with subsequent measurement of Potassium and Lactate Dehydrogenase (LDH), markers sensitive to hemolysis. Each sample was submitted to one of the following scenarios: (1) Centrifugation in a physical unit within two hours after collection; (2) Immediate centrifugation en route, using the E8F; and (3) Centrifugation in a physical unit after the end of the home route. In the second stage, the potassium results obtained from patients treated at home in the current (3) and idealized (2) scenarios were compared, in order to evaluate the direct impact of en route centrifugation on the quality of the tests. Results Potassium and LDH were the analytes that had the greatest impact with late centrifugation. Potassium presented a mean bias of 5.81% and Pearson's coefficient of 0.62. For the LDH, the mean bias was 6.37%, with Pearson of 0.83, when comparing scenarios 1 and 3. The results of the second phase evaluated with potassium showed that 15% of the patients had values impacted by late centrifugation after the route, changing the clinical judgment. Conclusion The validation of the E8 portable vehicule centrifuge demonstrated excellent performance and efficacy, resulting in improvements in the pre-analytical phase, reduction of hemolysis and recollections, reliability of results, increased team productivity, and improved patient experience.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

A-292 Innovation in Home Phlebotomy Services Using an In-Vehicle Centrifuge — Science Skim