Assessing the reliability and validity for metabolic syndrome diagnosis using point-of-care analyzers in a mobile setup - a pilot study.
Sanne Houtenbos, Anne Schraplau, Pia-Maria Wippert, Michael Rapp, Heinz Völler, Klaus Bonaventura, Frank Mayer
Medical Sociology and Psychobiology, Faculty of Human Sciences, University of Potsdam, Am Neuen Palais 10, Potsdam, 14469 , Germany. sanne.houtenbos@uni-potsdam.de. · Sports Medicine and Sports Orthopedics, University Outpatient Clinic, University of Potsdam, Potsdam, Germany.
Key points
From the paper's conclusion- A mobile setup using the Alere POC analyzer could be used as a valid mobile screening method for MetS to ensure early detection of risk factors.
Abstract
BACKGROUND: The metabolic syndrome (MetS) is characterized by having three or more of these risk factors: increased blood glucose (Glc), serum triglycerides (TG), blood pressure (BP), waist circumference (WC), and reduced high-density lipoprotein (HDL). Mobile diagnostics using point-of-care (POC) blood analyzers could help identify individuals at risk. This pilot study assessed reliability and validity of MetS risk factor analysis using POC analyzers in a mobile examination unit. METHODS: Fifty participants (52 ± 7y; 170 ± 10 cm; 80 ± 19 kg, m/f: 21/29) were included for analysis. The mobile examination unit was equipped with different POC analyzers (Alere, LABGEO, CardioChek for Glc, HDL, TG; Quo-Test-A1C Testkit for HbA1c'). Agreement compared to a reference laboratory (RefLab) was analyzed by Bland-Altman (bias, Limits of Agreement (LoA)), ICC, SEM and McNemar's test. Test-retest analysis was assessed on two different days (max. two weeks apart) using Spearman's rho, ICC, test-retest variability (TRV%) and McNemar's test. RESULTS: Based on the smallest bias and low number of missing values (Glc: -5 [LoA - 25 to 6] mg/dl; TG: 8 [LoA - 58 to 30] mg/dl; HDL: -2 [LoA - 14 to 9] mg/dl), and the highest agreement in MetS diagnosis compared to the results of RefLab, the Alere POC device was chosen. Test-retest reliability (n = 16) for MetS risk factors and HbA1c measured in the mobile setup ranged from poor to excellent with low TRV (0.9-5.8%). Risk factor classification and MetS diagnosis did not differ between measurement days. CONCLUSIONS: A mobile setup using the Alere POC analyzer could be used as a valid mobile screening method for MetS to ensure early detection of risk factors.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- A-362 Analytical validation of a semi-automated portable IoT-enabled POCT device for screening of non-communicable diseasesClinical Chemistry · 2026
- Analytical Agreement and Diagnostic Performance of a Commercial Point-of-Care Lipid Testing System for Lipid Profile and Atherogenic Ratios.Diagnostics (Basel, Switzerland) · 2026
- Point-of-Care Quantification of HbA1c and CRP using Finecare: Analytical Validation of a Fluorescence-Based System.Annals of clinical biochemistry · 2026
- Comparison of rapid portable devices and laboratory methods to assess the blood chemistry of farmed adult common snook (Centropomus undecimalis).Veterinary research communications · 2026
- Reliability and analytical validity of the MX3 hydration testing system during exercise in the heat.European journal of applied physiology · 2026
- B-030 Analytical Performance Evaluation of the Maccura Glycosylated Hemoglobin Analyzer G 02 (High-Performance Liquid Chromatography)Clinical Chemistry · 2026