Concordance of Corneal Tomographic Measurements Obtained via Swept-Source Optical Coherence Tomography Versus Scheimpflug Tomography.
Kamini Narendra Reddy, Ramadhan Ahmed, Matthew Zeng, Nakul S Shekhawat
Wilmer Eye Institute, Johns Hopkins University School of Medicine, Baltimore, MD 21287, USA. · School of Medicine, Johns Hopkins University School of Medicine, Baltimore, MD 21205, USA.
Key points
From the paper's conclusion- Anterion and Pentacam HR showed satisfactory-to-excellent measurement concordance, with negligible mean inter-device biases for most anterior keratometry measurements and anterior astigmatism vectors. However, individual-eye differences could approach 0.49-0.61 D for anterior keratometry and exceed 20 μm for pachymetry. These differences may become relevant in borderline cases and during longitudinal disease monitoring, warranting parameter-specific caution.
Abstract
PURPOSE: To compare concordance of corneal tomographic measurements obtained using Anterion versus Pentacam HR. SETTING: Outpatient ophthalmology clinic. DESIGN: Retrospective cross-sectional study. METHODS: Eyes without corneal abnormalities underwent tomography using the Anterion Cornea App and Pentacam HR. Measurements included anterior keratometry (K1, K2, ΔK, average K, Kmax), posterior keratometry (K1, K2, ΔK, and average K), astigmatism power-vector components (J0, J45) for both surfaces, central corneal thickness (CCT), white-to-white (WTW), and thinnest pachymetric point (TP). We assessed agreement using Pearson correlation coefficients (ρ), Lin's concordance correlation coefficient (CCC), and Bland-Altman analysis. Generalized estimating equation models were used to assess measurement differences between devices. RESULTS: A total of 64 eyes of 41 patients underwent corneal tomography with both devices. Anterion and Pentacam HR showed strong linear correlations for all measurements (ρ > 0.700), with concordance ranging from satisfactory to excellent (CCC 0.704-0.981). No statistically significant mean inter-device differences were found for anterior K1 (0.01 diopters [D], p = 0.775), K2 (-0.003 D, p = 0.951), average K (0.01 D, p = 0.801), ΔK (-0.02 D, p = 0.693), or the anterior astigmatism vectors J0 (0.002 D, p = 0.936) and J45 (0.03 D, p = 0.129). Mean posterior K1, K2, and average K measurements were flatter with Anterion than Pentacam HR, with mean inter-device differences of 0.13 D (p < 0.001), 0.11 D (p < 0.001), and 0.12 D (p < 0.001), respectively. The mean inter-device difference for posterior ΔK was -0.02 D (p = 0.018). Statistically significant mean inter-device differences were also found for posterior J0 (-0.01 D, p = 0.018) and J45 (0.01 D, p = 0.007). Anterion measured lower values than Pentacam HR for anterior Kmax, CCT, and TP, with mean inter-device differences of -0.24 D (p = 0.002), -7.66 μm (p < 0.001), and -8.58 μm (p < 0.001), respectively, and a higher mean WTW, with a mean inter-device difference of 0.07 mm (p < 0.001). CONCLUSIONS: Anterion and Pentacam HR showed satisfactory-to-excellent measurement concordance, with negligible mean inter-device biases for most anterior keratometry measurements and anterior astigmatism vectors. However, individual-eye differences could approach 0.49-0.61 D for anterior keratometry and exceed 20 μm for pachymetry. These differences may become relevant in borderline cases and during longitudinal disease monitoring, warranting parameter-specific caution.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Measurement Repeatability and Concordance of Corneal Tomography in Keratoconus Using Anterion Versus Pentacam.Bioengineering (Basel, Switzerland) · 2026
- Concordance of Cataract Biometry Measurements Obtained via Swept-Source Optical Coherence Tomography Versus Optical Biometry.Bioengineering (Basel, Switzerland) · 2026
- Agreement between Sirius Scheimpflug-Placido tomography and Anterion swept-source OCT in keratoconic eyes stratified by prior corneal cross-linking: a retrospective cross-sectional agreement study.International ophthalmology · 2026
- Quantitative assessment of corneal haze: comparative optical densitometry using anterior segment OCT and Pentacam.International ophthalmology · 2026
- Comparison of anterior segment parameters in cataract patients with different axial lengths using IOL Master and Pentacam.BMC ophthalmology · 2026
- Comparison of Anterior Segment Biometric Parameters Measured by Pentacam and Sirius in Healthy Myopic Eyes.Photodiagnosis and photodynamic therapy · 2026