B-137 Diazyme and Freelite Serum Free Light Chain Assays Are Not Clinically Interchangeable at IMWG-Defined Decision Points

Stuart Morgan, Sanchia Sarin, Charlotte Crawford, Jake Fisher, Hannah Jones, Luke Moore, Syreeta Allen, Jamieleigh Logan,

Key points

From the paper's conclusion
  • Conclusion This study demonstrates that Diazyme and Freelite FLC assays are not analytically or clinically interchangeable.

Abstract

Background Serum free light chain (sFLC) measurements are integral to the diagnosis and monitoring of Monoclonal Gammopathies. The International Myeloma Working Group (IMWG) diagnostic criteria for sFLC testing in Multiple Myeloma were established using Freelite® assays. Given the availability of alternative sFLC assays, including Diazyme FLC assays, evaluation of analytical agreement and clinical concordance at IMWG-defined medical decision points is essential to support guideline-consistent interpretation. Methods An analytical comparison was conducted using Diazyme FLC assays (Diazyme Laboratories) on the Roche cobas™ c503 analyzer (Roche Diagnostics) and Freelite sFLC assays on the Optilite® analyser (The Binding Site). A total of 125 serum samples were included, comprising healthy donor samples and clinical samples spanning a range of monoclonal gammopathies, including Multiple Myeloma, Smoldering Multiple Myeloma, AL Amyloidosis, and Waldenström’s Macroglobulinemia. Quantitative agreement between assays was assessed using Passing-Bablok regression and coefficients of determination (R2), and systematic bias was evaluated using Bland-Altman analysis. Result reportability was assessed relative to assay limits of quantitation (LoQ). Clinical concordance was evaluated at IMWG-defined sFLC medical decision points, including the Myeloma-defining SLiM criterion (involved/uninvolved sFLC ratio of at least 100 with involved sFLC concentration of at least 100 mg/L) and the Smoldering Multiple Myeloma risk stratification criterion (involved/uninvolved sFLC ratio >20), with Freelite assays treated as the clinical reference. Results Freelite assays generated reportable kappa/lambda FLC ratios for 100% of clinical samples, compared with 87.7% using Diazyme assays, reflecting lower LoQ and greater analytical sensitivity. Quantitative comparison demonstrated systematic analytical differences between assays. Passing-Bablok regression demonstrated poor analytical agreement between Diazyme and Freelite assays across the measuring range. R2 values were below the commonly used threshold of 0.95 for interchangeability (kappa-FLC 0.654, lambda-FLC 0.485, kappa/lambda ratio 0.857), with evidence of proportional positive bias for lambda-FLC and negative bias for kappa/lambda ratio measurements. Bland-Altman analysis of clinical samples showed substantial mean bias, including -19.8% for kappa-FLC, +51.9% for lambda-FLC, and -67.5% for kappa/lambda ratios with Diazyme relative to Freelite assays. Clinically meaningful discordance was observed at IMWG medical decision points. Among samples classified as SLiM-positive by Freelite assays, Diazyme assays reproduced the same classification in only 42.9% of cases; 42.9% were classified as SLiM-negative and 14.3% yielded non-calculable ratios due to results below the assay LoQ. At the Smoldering Multiple Myeloma risk stratification threshold, Diazyme assays reproduced Freelite-defined risk classification in only 42.3% of cases. In a representative diagnostic case, a Freelite assay result fulfilled the IMWG SLiM Myeloma-defining criterion and aligned with the final Multiple Myeloma diagnosis, whereas Diazyme assay results did not meet the SLiM threshold due to a higher uninvolved sFLC value, resulting in discordant classification. Conclusion This study demonstrates that Diazyme and Freelite FLC assays are not analytically or clinically interchangeable. Differences in analytical sensitivity, systematic bias, and result reportability lead to clinically meaningful discordance at IMWG-defined FLC medical decision points. These findings highlight the importance of assay selection when interpreting FLC results relative to established IMWG guideline criteria to support consistent interpretation of FLC results in clinical practice.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers