Clinical implications of lateral head mis-centering on eye lens dose in pediatric computed tomography examinations: a phantom study.
Mohammad Aljamal, Zuhal Y Hamd, Awadia Gareeballah, Mousa Shaheen, Lama Almudaimeegh, Amna M Ahmed, Mohammed Alharbi, Abbas Abuthabet,
Faculty of Allied Medical Sciences, Department of Medical Imaging, Arab American University, Jenin, Palestine. · Department of Radiological Sciences, College of Health and Rehabilitation Sciences, Princess Nourah bint Abdulrahman University, Riyadh, Saudi Arabia.
Key points
From the paper's conclusion- Lateral mis-centering introduces systematic dose distribution asymmetry and significant discrepancies in CT dose reporting. These findings highlight the importance of accurate patient positioning and caution in interpreting system-reported dose metrics in pediatric imaging.
Abstract
BACKGROUND AND AIM: Patient mis-centering during pediatric CT examinations can significantly affect radiation dose distribution, particularly to radiosensitive organs such as the eye lens. This study quantitatively evaluates the effect of lateral head mis-centering on eye lens dose and investigates discrepancies between measured and system-reported dose metrics using a low-cost pediatric head phantom. MATERIALS AND METHODS: A low-cost tissue-equivalent pediatric head phantom was constructed locally and validated using CT number comparisons with reference tissue values. Eye lens dose was measured using a pencil ionization chamber at the isocenter and at ±1-3 cm lateral displacements on two CT systems. Measured dose length product (DLP) values were compared with CT-reported DLP and air kerma-derived estimates. The data was analyzed statistically using Pearson correlation, paired sample t-tests, and one-way ANOVA. RESULTS: Strong correlations were observed between isocenter and off-center doses (r = 0.994-0.998, p < 0.001). Although positional differences were not statistically significant (p > 0.05), a consistent asymmetric trend was observed, with higher doses on the right side. System-reported DLP substantially overestimated measured values (e.g., 720 vs. 304.83 mGy·cm). CONCLUSION: Lateral mis-centering introduces systematic dose distribution asymmetry and significant discrepancies in CT dose reporting. These findings highlight the importance of accurate patient positioning and caution in interpreting system-reported dose metrics in pediatric imaging.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Vertical miscentering increases eye lens radiation dose and degrades image quality in pediatric head CT: insights from an in-house heterogeneous phantom.Neuroradiology · 2026
- Comparison of ocular radiation doses between ultra-low-dose photon-counting computed tomography and digital radiography for orbital foreign body screening: A phantom study.Radiography (London, England : 1995) · 2026
- Influence of phantom size on the IEC 60601-2-44 wide-beam CTDI correction: A Monte Carlo studyEPJ Web of Conferences · 2026
- Calculation of the conversion coefficient from kerma to eye lens dose for computerized tomography head phantom.Radiation protection dosimetry · 2026
- Patient Centring in CT: An International Survey of Clinical Practice and Future Directions.Journal of medical radiation sciences · 2026
- Localiser-Based Software-Assisted Patient Positioning in Chest CT: A Propensity Score-Matched Study.Diagnostics (Basel, Switzerland) · 2026