Influence of phantom size on the IEC 60601-2-44 wide-beam CTDI correction: A Monte Carlo study

Zouhair Airouss, Yacine El Attaoui, Lhoucine Oufni

Key points

From the paper's conclusion
  • We investigated the dependence of the resulting correction on polymethyl methacrylate (PMMA) phantom diameter using a GATE v9.1/Geant4 v10.7 model of a Siemens SOMATOM Emotion scanner.

Abstract

A 100 mm pencil ionization chamber cannot encompass the complete longitudinal dose profile of wide computed-tomography (CT) beams. DEC 60601-2-44 therefore combines a narrow-beam, inphantom CTDI reference with a free-in-air wide-to-reference dose ratio. We investigated the dependence of the resulting correction on polymethyl methacrylate (PMMA) phantom diameter using a GATE v9.1/Geant4 v10.7 model of a Siemens SOMATOM Emotion scanner. The model was compared with CTDI w measurements in 16 and 32 cm phantoms at 10 mm collimation and 80, 110 and 130 kVp; deviations calculated from reported rounded doses were 3.32–7.23%. Nominal collimations of 40–200 mm were subsequently studied by numerical extension of the beam aperture. The uncorrected efficiencies decreased to approximately 0.50 (16 cm) and 0.45 (32 cm) at 200 mm, whereas DEC-corrected values remained near 0.84 and 0.76. Across nine PMMA diameters (8–40 cm), the mean corrected efficiency decreased from 0.889 to 0.728, with a maximum within-diameter spread of 0.015 across beam widths. A first-order uncertainty bound derived from the reported per-cell Monte Carlo precision is presented to contextualize the observed variations. The corresponding model-specific conversion factor to the 900 mm integrated dose index increased from 1.125 to 1.374. These results show that the DEC method largely compensates for collimation while retaining an object-size dependence in homogeneous PMMA geometry.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers