Influence of phantom size on the IEC 60601-2-44 wide-beam CTDI correction: A Monte Carlo study
Zouhair Airouss, Yacine El Attaoui, Lhoucine Oufni
Key points
From the paper's conclusion- We investigated the dependence of the resulting correction on polymethyl methacrylate (PMMA) phantom diameter using a GATE v9.1/Geant4 v10.7 model of a Siemens SOMATOM Emotion scanner.
Abstract
A 100 mm pencil ionization chamber cannot encompass the complete longitudinal dose profile of wide computed-tomography (CT) beams. DEC 60601-2-44 therefore combines a narrow-beam, inphantom CTDI reference with a free-in-air wide-to-reference dose ratio. We investigated the dependence of the resulting correction on polymethyl methacrylate (PMMA) phantom diameter using a GATE v9.1/Geant4 v10.7 model of a Siemens SOMATOM Emotion scanner. The model was compared with CTDI w measurements in 16 and 32 cm phantoms at 10 mm collimation and 80, 110 and 130 kVp; deviations calculated from reported rounded doses were 3.32–7.23%. Nominal collimations of 40–200 mm were subsequently studied by numerical extension of the beam aperture. The uncorrected efficiencies decreased to approximately 0.50 (16 cm) and 0.45 (32 cm) at 200 mm, whereas DEC-corrected values remained near 0.84 and 0.76. Across nine PMMA diameters (8–40 cm), the mean corrected efficiency decreased from 0.889 to 0.728, with a maximum within-diameter spread of 0.015 across beam widths. A first-order uncertainty bound derived from the reported per-cell Monte Carlo precision is presented to contextualize the observed variations. The corresponding model-specific conversion factor to the 900 mm integrated dose index increased from 1.125 to 1.374. These results show that the DEC method largely compensates for collimation while retaining an object-size dependence in homogeneous PMMA geometry.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Influence of gantry overrun angle on CTDI variability: a Monte Carlo study.Radiological physics and technology · 2026
- Development of a slice‐specific ct dose library with interpolation methods for organ dosimetry in MIRDctJournal of Applied Clinical Medical Physics · 2026
- Calculation of the conversion coefficient from kerma to eye lens dose for computerized tomography head phantom.Radiation protection dosimetry · 2026
- Clinical implications of lateral head mis-centering on eye lens dose in pediatric computed tomography examinations: a phantom study.Frontiers in medicine · 2026
- Survey of normalized CTDI vol values across four major computed tomography vendors for use in the MIRDct softwareJournal of Applied Clinical Medical Physics · 2026
- Evaluation of a fast kV switching dual energy CT in deriving relative stopping power using tissue equivalent phantoms of various sizesMedical Physics · 2026