Localiser-Based Software-Assisted Patient Positioning in Chest CT: A Propensity Score-Matched Study.
Xinnan Shen, Li Tao, Wen Li, Youhong Cao, Jianping Liu, Yiting Wang, Meiqi Wang, Zhiwei Zhang
Department of Radiology, The First Affiliated Hospital of Chongqing Medical University, No. 1 Youyi Road, Yuzhong District, Chongqing 400016, China.
Key points
From the paper's conclusion- Methods: In this retrospective, single-centre study, 192 propensity score-matched pairs from 423 adults undergoing non-contrast chest CT were analysed.
Abstract
Background/Objectives: Vertical off-centring in chest CT impairs bowtie-filter beam-hardening correction, degrading noise uniformity and image quality. iCentering derives the patient's geometric centre from a single posteroanterior localiser and automatically adjusts table height. This study aimed to evaluate whether iCentering reduces off-centring distance (OCD) and improves image quality versus manual positioning in chest CT, and to investigate correlations between OCD and noise characteristics. Methods: In this retrospective, single-centre study, 192 propensity score-matched pairs from 423 adults undergoing non-contrast chest CT were analysed. OCD, objective image quality (standard deviation (SD), signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), figure of merit (FOM)), noise distribution ratio (NDRA-P), and radiation dose indices were measured. Paired comparisons employed t-tests or Wilcoxon signed-rank tests with 95% confidence intervals (CIs) and effect sizes reported. Directionally stratified Spearman correlation and multivariable regression examined OCD-noise associations. Results: Intra- and inter-observer reliability was good to excellent (ICC 0.87-0.99 for OCD; 0.75-0.97 for most regions of interest). iCentering moved median OCD closer to zero at all levels (Level 1: -7.25 vs. -29.40 mm; Level 2: -1.45 vs. -22.00 mm; Level 3: 7.45 vs. -11.90 mm; all p < 0.001; r range: 0.76-0.77). Regarding image quality, SD decreased and SNR, CNR, and FOM increased at all levels (all p < 0.001), while NDRA-P approached 1.00 (p < 0.05). Radiation dose indices were unchanged (p > 0.05). OCD correlated with SD (|ρ| = 0.478-0.812, p < 0.001); regression confirmed independent associations in both strata (OCD < 0: β range: -0.068 to -0.079, R2 = 0.52-0.65; OCD > 0: β range: 0.056-0.078, R2 = 0.28-0.54), whereas the association between OCD and NDRA-P was weak. Conclusions: iCentering improves centring accuracy and objective image quality in chest CT, without increasing radiation exposure. Reduced OCD is independently associated with lower image noise, underscoring the value of precise centring.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Impact of patient centering on radiation dose and image noise in computed tomography: Noise reduction potential of adaptive statistical iterative reconstructionEPJ Web of Conferences · 2026
- Patient Centring in CT: An International Survey of Clinical Practice and Future Directions.Journal of medical radiation sciences · 2026
- From CT Centring Knowledge to Verified Benefit in Routine Practice.Journal of medical radiation sciences · 2026
- Response to "From CT Centring Knowledge to Verified Benefit in Routine Practice".Journal of medical radiation sciences · 2026
- Automated patient-derived noise power spectrum estimation in abdominal CT: a preliminary validationBiomedical Physics & Engineering Express · 2026
- Vertical miscentering increases eye lens radiation dose and degrades image quality in pediatric head CT: insights from an in-house heterogeneous phantom.Neuroradiology · 2026