Radiation protection dosimetry

Calculation of the conversion coefficient from kerma to eye lens dose for computerized tomography head phantom.

Natasa Danilovic, Jelena Stankovic Petrovic, Predrag Bozovic, Dusan Topalovic, Ivana Pepic, Natasa Todorovic, Jovana Nikolov

University of Novi Sad, Faculty of Sciences, Department of Physics, Trg Dositeja Obradovica 4, 21000 Novi Sad, Serbia. · Department of Radiation and Environmental Protection, Vinca Institute of Nuclear Sciences, National Institute of the Republic of Serbia, University of Belgrade, P.O. box 522, 11001 Belgrade, Serbia.

Key points

From the paper's conclusion
  • This study aims to calculate Hp(3) CCs for the computerized tomography (CT) phantom (16 cm diameter, 15 cm height) and compare them with CCs for head phantom (20 × 20 cm) given in the Part 3 of international standard ISO 4037 (ISO 4037-3) to improve eye lens dosimetry methodologies in Serbia.

Abstract

Personal dosemeters calibration relies on the air kerma conversion coefficients (CCs), which depends on the phantom and used exposure geometry. This study aims to calculate Hp(3) CCs for the computerized tomography (CT) phantom (16 cm diameter, 15 cm height) and compare them with CCs for head phantom (20 × 20 cm) given in the Part 3 of international standard ISO 4037 (ISO 4037-3) to improve eye lens dosimetry methodologies in Serbia. Monte Carlo radiation transport simulations using MCNP Version 6.1.0 were performed for S-Cs (662 keV) and S-Co (1.25 MeV) reference photon fields. Absorbed dose and air kerma values were derived to calculate Hp(3) CCs for both CT and ISO phantoms across all specified angles for two radiation qualities. Calculated CCs were compared against the CCs ISO phantom values and previously published data to evaluate the relative deviations. This work provides the necessary foundation to determine whether correction coefficients should be introduced during the calibration and type-testing of passive eye dosimetry systems. Highlights MCNP 6.1.0 was used to derive conversion coefficients for a computerized tomography (CT) head phantom.The CT phantom caused backscatter deficits in S-Cs and S-Co photon fields.Posterior irradiation led to up to 20% overestimation for S-Cs.Dimension-specific corrections are needed for angular response testing.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers