Bridging Gaps in Newborn Screening: A Comparative Study of Point-of-Care Testing Versus Traditional Reference Laboratory Method for Congenital Hypothyroidism.
Endah Indriastuti, Rahmah Yasinta Rangkuti, Fatimah Nur Fitriani, Putri Alief Siswanto, Edwin Nugroho Njoto, Muhammad Nazhif Haykal, Fauqa Arinil Aulia
Department of Medicine, Faculty of Medicine and Health, Institut Teknologi Sepuluh Nopember, Surabaya, Indonesia. · Department of Medical Profession, Faculty of Medicine and Health, Institut Teknologi Sepuluh Nopember, Surabaya, Indonesia.
Key points
From the paper's conclusion- POCT devices present a promising alternative to traditional reference screening method for congenital hypothyroidism.
Abstract
OBJECTIVE: This study aimed to compare the reliability, agreement, and practical application (turnaround time) of a point-of-care testing (POCT) device using two different reagents versus the traditional reference laboratory method for thyroid-stimulating hormone (TSH) measurement in newborn congenital hypothyroidism screening. METHODOLOGY: This prospective cross-sectional method-comparison study included 122 term newborns at a mother and child hospital in Surabaya, Indonesia. Heel-prick capillary blood samples were simultaneously analyzed using the Standard F-200 POCT device with two reagents (Group 1: n = 62, Standard™ F TSH-II; Group 2: n = 60, FastClear F TSHII) and compared to the standard dried blood spot method (GSP® Neonatal hTSH, PerkinElmer) as reference. Analyses included Pearson and Spearman correlations, linear regression, Bland-Altman plots, intraclass correlation coefficient (ICC), Cohen's Kappa, and mean difference tests. RESULTS: Strong positive correlations were observed in both Group 1 (Pearson's r = 0.87, Spearman's ρ = 0.85; p <0.001) and Group 2 (Pearson's r = 0.88, Spearman's ρ = 0.91; p <0.001). Bland-Altman analysis showed minimal systematic bias with acceptable agreement between the two techniques. The ICC values indicated excellent reliability (Group 1: 0.89; Group 2: 0.91). High categorical agreement was confirmed by Cohen's Kappa values of 0.85 and 0.87 (p <0.001). No significant differences in mean TSH levels were found between the POCT and reference methods. The POCT turnaround time was 30 minutes, significantly shorter than the 15-22 days required for the traditional method. CONCLUSION: POCT devices present a promising alternative to traditional reference screening method for congenital hypothyroidism.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Evaluation of point-of-care complete blood count and C-reactive protein testing for routine use in a pediatric emergency department.European journal of pediatrics · 2026
- Screen Positivity and Feasibility of Newborn Thyroid Screening Using Postnatal Venous Thyroid-Stimulating Hormone: A Retrospective Cohort Study From Northeast India.Cureus · 2026
- Incidence and clinical distribution of congenital hypothyroidism in Bolu Province, Türkiye: a four-year analysis of the National Newborn Screening Program.Current medical research and opinion · 2026
- Validation of Point-of-care Testing Device HemoCue® Glucose 201RT in Blood Glucose Estimation.Indian journal of community medicine : official publication of Indian Association of Preventive & Social Medicine · 2026
- A-148 Decentralized TSH Testing Using Quantitative Capillary Microsampling: Analytical and Clinical Agreement of Capitainer®B50 with Venous PlasmaClinical Chemistry · 2026
- Comparative Performance Analysis of Three Point-of-Care Glucose Testing Devices against a Central Laboratory Reference Method.EJIFCC · 2026