Clinical Chemistry

A-148 Decentralized TSH Testing Using Quantitative Capillary Microsampling: Analytical and Clinical Agreement of Capitainer®B50 with Venous Plasma

Mathias Karlsson, Anders Larsson, Marc Albeek, Johan Olausson, Helena Filipsson Nyström, Monica Lindell, Agneta Lindö

Key points

From the paper's conclusion
  • We evaluated the analytical performance and clinical agreement of TSH using the Capitainer®B50 quantitative dried blood microsampling device compared with venous plasma.

Abstract

Background Thyroid stimulating hormone (TSH) is the cornerstone biomarker for diagnosing and monitoring thyroid dysfunction, particularly hypothyroidism, which requires long term follow up and frequent testing. Venous sampling may create logistical barriers for patients and healthcare systems. Quantitative microsampling could support decentralized and patient centred care by enabling reliable home based capillary sampling. We evaluated the analytical performance and clinical agreement of TSH using the Capitainer®B50 quantitative dried blood microsampling device compared with venous plasma. Methods Adults undergoing routine TSH testing self collected capillary blood using Capitainer®B50 devices following standardized video and written instructions. Venous lithium heparin plasma was collected at the same visit. Microsamples were mailed to the analysis laboratory and processed on a Cobas® e411 platform; venous samples were analyzed on Cobas® e411 or Cobas® pro e801. Agreement between sample types was assessed using linear regression, Passing–Bablok, Deming regression, and Bland–Altman analyses. Analytical precision for duplicate microsamples was evaluated using coefficients of variation (CV). Stability was tested at room temperature and under simulated hot and cold shipping for up to 10 days. Results Seventy-nine participants were included; 76 successfully obtained at least one valid microsample. The median CV for duplicate microsamples was 2.5% (IQR 0.5–4.9%), and no CV exceeded 20%. Strong correlation was observed between microsample derived and plasma TSH values (R² ˜ 0.99). After calibration, microsampling yielded plasma equivalent results with minimal systematic bias. Bland–Altman analysis showed no proportional bias, although imprecision increased at the lowest TSH concentrations. Stability testing demonstrated no degradation across all storage conditions. Conclusion Quantitative capillary microsampling using Capitainer®B50 provides clinically reliable, plasma equivalent TSH measurements with high precision. The method is well suited to decentralized monitoring, supports integrated and patient centered care models, and may reduce the need for venous sampling, particularly for patients requiring frequent long term TSH follow up.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers