Evaluation of point-of-care complete blood count and C-reactive protein testing for routine use in a pediatric emergency department.
Giancarlo Lai, Antonio Torchia, Gioconda Aiello, Domitilla Parigi, Alessandro Campagnoli, Paola Tommasi, Gianvincenzo Zuccotti, Felicia Stefania Falvella,
Clinical Pathology Unit, ASST Fatebenefratelli-Sacco, Milan, Italy. · Pediatric Department, Buzzi Children's Hospital, ASST Fatebenefratelli-Sacco, Milan, Italy.
Key points
From the paper's conclusion- Cautious interpretation is warranted for WBC, and particularly for RBC parameters and PLT. The observed data dispersion may reflect pre-analytical variability, potentially affecting the precision of the findings and the applicability of POCT in routine emergency practice, especially for capillary samples.
Abstract
BACKGROUND: Point-of-care testing (POCT) of complete blood count (CBC) and C-reactive protein (CRP) may support rapid decision-making in pediatric emergency care. We evaluated the performance of the Horiba Microsemi analyzer operated by clinical staff in the emergency department (ED) of V. Buzzi Children's Hospital (Milan, Italy) compared with laboratory testing. METHODS: An unselected cohort of 195 patients (108 males and 87 females; median age 6 years, IQR 2-11 years) admitted to the ED was enrolled over 6 weeks to reflect routine clinical practice. Capillary and venous samples were analyzed using the Horiba Microsemi CRP LC-767G and compared with simultaneously collected venous samples analyzed by laboratory reference systems (Sysmex XN-9000 for CBC and Abbott Alinity for CRP). Method comparison was performed using MedCalc Statistical Software, applying Passing-Bablok regression and Bland-Altman analysis. Accuracy was assessed by comparing mean bias with European Federation of Clinical Chemistry and Laboratory Medicine (EFLM) biological variation-based performance specifications. RESULTS: Strong correlation (Spearman correlation coefficient, ρ ≥ 0.90) between POCT (capillary and venous) and laboratory measurements was observed for mean corpuscular volume (MCV), mean corpuscular hemoglobin (MCH), and C-reactive protein (CRP). For lymphocyte (LYMPH) and granulocyte (GRAN) counts, a strong correlation was observed only in venous samples. Moderate correlation (0.75 ≤ ρ < 0.90) was observed for white blood cells (WBC) and platelets (PLT) in both matrices, for hemoglobin (HGB) in venous samples only, and for lymphocytes (LYMPH) and granulocytes (GRAN) in capillary samples. Poor correlation (ρ < 0.75) was found for red blood cells (RBC), hematocrit (HCT), mean corpuscular hemoglobin concentration (MCHC), and monocytes (MONO) in both matrices, and for HGB in capillary samples. According to EFLM specifications, MCHC, LYMPH, and CRP met the predefined desirable performance specifications in both matrices, whereas WBC and GRAN met these criteria only in venous POCT samples. Significant bias was observed for RBC parameters and PLT. CONCLUSION: Cautious interpretation is warranted for WBC, and particularly for RBC parameters and PLT. The observed data dispersion may reflect pre-analytical variability, potentially affecting the precision of the findings and the applicability of POCT in routine emergency practice, especially for capillary samples.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Implementation and staff perceptions of point-of-care full blood count and C-reactive protein testing in a UK tertiary paediatric emergency department.European journal of pediatrics · 2026
- Point-of-Care White Blood Cell Differential Testing in Pediatric Inpatients (2 Months-17 Years): How Does HemoScreen Perform?International journal of laboratory hematology · 2026
- Point-of-Care Quantification of HbA1c and CRP using Finecare: Analytical Validation of a Fluorescence-Based System.Annals of clinical biochemistry · 2026
- Comparison of the HemoScreen Point-Of-Care Haematology Analyser With Sysmex XN, Beckman Coulter DxH 560 and HORIBA Yumizen H550 in Patients With Abnormal Blood Counts.International journal of laboratory hematology · 2026
- Analytical validation of capillary blood Krebs von den Lungen-6 measurement using a high- sensitivity chemiluminescent immunoassay in pediatric patients.Frontiers in medicine · 2026
- Capillary Blood Collection Methods: An assessment of different collections technologies compared to traditional venous sampling.Annals of clinical biochemistry · 2026