The Clinical Biochemist Reviews

Implementation of high-sensitivity troponin point of care testing to support rapid chest pain assessment across Western Australia

Elizabeth A. Byrnes, Sasa Trivic, Leya A. Grego, Eric A. Magtuto, Roxanna Stevens, Zoe D. Benham, Ee Mun Lim, Rachel J.M. Dennis

Key points

From the paper's conclusion
  • The implementation of the new hsTnI to replace existing cardiac TnI testing across WA was a large project, given the extensive regional area of the state, an area of 2.5 million square kilometres.

Abstract

Point of care (POC) testing is an important part of frontline health care in Western Australia (WA), with testing performed by clinical staff in a variety of urgent care settings. Cardiac troponin testing on POC devices has been available for several years, however the assays on POC devices were not considered high sensitivity. In 2025, the PathWest POC department evaluated new to market high sensitivity troponin I (hsTnI) tests on four POC devices, and proceeded with a roll-out out of hsTnI on the existing fleet of Abbott iSTAT instruments. The implementation of the new hsTnI to replace existing cardiac TnI testing across WA was a large project, given the extensive regional area of the state, an area of 2.5 million square kilometres. Evaluation of the iSTAT hsTnI cartridge was performed, including interference testing, and a managed roll-out of the test across 130 ISTAT instruments in WA was undertaken.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.