Diagnostic accuracy of high-sensitivity point-of-care troponin assays in emergency department for patients presenting with chest pain: a systematic review.
Lara E van Daalen, Bart L F Schueler, Paul O Collinson, Barbra E Backus
University of Amsterdam, Amsterdam. · Emergency Department, Dijklander Hospital, Hoorn, The Netherlands.
Key points
From the paper's conclusion- Hs-cTn POC assays are reliable for integration into the diagnostic pathways for patients presenting with chest pain at the ED, mainly when used for rapid rule-out of AMI in patients with very low troponin values upon arrival.
Abstract
High-sensitivity cardiac troponin (hs-cTn) assays have significantly enhanced the diagnosis in patients suspected of acute coronary syndrome (ACS). New hs-cTn point-of-care (POC) devices (turnaround time <20 min) may significantly reduce time to diagnosis. We evaluated the diagnostic accuracy of four currently commercially available hs-cTn POC devices to assess their potential for clinical implementation at the emergency department (ED). We conducted a search in PubMed Central on 20 November, 2025. Studies were included if they investigated patients greater than or equal to 18 years presenting with chest pain suggestive of ACS and evaluated commercially available hs-cTn POC devices. Studies had to report diagnostic accuracy [sensitivity, specificity, negative predictive value (NPV), positive predictive value] for the diagnosis of acute myocardial infarction (AMI). Studies were assessed for risk of bias and applicability using the QUADAS-3 tool. We identified 436 articles. The 11 included studies consisted of both derivation and validation studies, including a total of 10 115 patients. The hs-cTn POC assays demonstrated the best diagnostic accuracy when applied as a single-test rule-out upon arrival at the ED, showing the following sensitivity and NPV: Atellica VTLi sensitivity 98.8-100% for troponin less than 4 ng/L (NPV: 99.8-100%) and 100% for troponin less than 7 ng/L (NPV 100%). Pathfast sensitivity 96.3-100% for troponin less than 3 ng/L (NPV: 98.1-100%). i-STAT sensitivity of 99.8% for troponin less than 5 ng/L (NPV: 100%). TriageTrue sensitivity of 100% for troponin less than 3 ng/L (NPV 100%). VTLi, Pathfast, and TriageTrue were also tested using the 99th percentile with a 0 + 1/2/3 h troponin protocol. Reported sensitivities: VTLi 64.8-100%, Pathfast 87% for plasma and 83% for whole blood, and TriageTrue 79.8%. Hs-cTn POC assays are reliable for integration into the diagnostic pathways for patients presenting with chest pain at the ED, mainly when used for rapid rule-out of AMI in patients with very low troponin values upon arrival.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- What is the diagnostic accuracy of different high sensitive point-of-care troponin assays for the assessments of patients presenting with chest pain at the emergency department?European Heart Journal · 2026
- A-041 Clinical Performance and Risk Classification Agreement of Point-of-Care i-STAT High-Sensitivity Troponin I Compared with Abbott Alinity hs-cTnI in ED Patients with Suspected Acute Coronary SyndromeClinical Chemistry · 2026
- Point-of-care high-sensitivity troponin testing for rapid chest pain triage in emergency primary care: preliminary efficacy findings from the OUT-POC StudyEuropean Heart Journal · 2026
- Impact of point-of-care high-sensitivity troponin testing on index admission and 30-day costs: secondary analysis of a randomised controlled trialEuropean Heart Journal · 2026
- ICare-FASTER (Improving Care by Faster Risk Stratification Through Use of High-Sensitivity Point-of-Care Troponin in Patients Presenting With Possible Acute Coronary Syndrome in the Emergency Department): A Stepped-Wedge Cluster-Randomized Trial.Circulation · 2026
- A 0/2-Hour Algorithm Using a Novel Point-of-Care High-Sensitivity Cardiac Troponin I Assay for Suspected Myocardial Infarction.American heart journal · 2026