Finite element analysis of a novel repair device for lumbar spondylolysis.
Bin Xie, Hongda Xu, Haitao Deng, Mingfan Li, Shengxin Zhao, Youpeng Hu, Li Liu, Peidong Qing
Department of Spine Surgery, Mianyang Orthopedic Hospital, Mianyang, Sichuan Province, China. · Department of Preventive Medicine, Sichuan College of Traditional Chinese Medicine, Mianyang, Sichuan Province, China.
Key points
From the paper's conclusion- The new type of lumbar pars interarticularis repair device is highly reliable in fixation, and the bone grafting area of the pars interarticularis is well exposed, providing favorable conditions for bone grafting and subsequent bone formation. It is a highly promising internal fixation device that offers a new option for pars interarticularis repair.
Abstract
BACKGROUND: This study aimed to compare the biomechanical differences between a novel repair device for lumbar spondylolysis and a nail-hook internal fixation system through three-dimensional finite element analysis. METHODS: Three-dimensional images of L4-S1 were constructed based on computed tomography images of a young male patient. Four finite element models were developed: Model A: intact model; Model B: bilateral spondylolysis of the L5 model; Model C: spondylolysis fixed by the nail-hook internal fixation system model; and Model D: spondylolysis fixed by the novel repair device model. A uniformly distributed 600-N vertical downward concentrated force and 10 Nm of torque were applied to the upper surface of the L4 to simulate six physiological activities of the lumbar spine. The fixation effects were evaluated by comparing the range of motion, maximum stress of the intervertebral disks, maximum displacement of the spondylolysis, and stress distribution of the internal fixation. RESULTS: The range of motions of the two internal fixation models were similar in each segment and direction. In terms of the maximum displacement of the spondylolysis, that of the fracture surface in the novel repair device model was significantly reduced compared with the nail-hook internal fixation model in each range of motion. The maximum stress of the intervertebral disks in the two models had their respective advantages in different activity directions, but the difference in stress values was not significant. The stress levels of both internal fixation devices were concentrated on the connecting rod and the junction of the pedicle screw and the connecting rod. Due to the shorter connecting rod of the nail-hook internal fixation system, the stress on the connecting rod was more concentrated, while the stress distribution on the connecting rod of the novel device was more dispersed. CONCLUSION: The new type of lumbar pars interarticularis repair device is highly reliable in fixation, and the bone grafting area of the pars interarticularis is well exposed, providing favorable conditions for bone grafting and subsequent bone formation. It is a highly promising internal fixation device that offers a new option for pars interarticularis repair.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Three-dimensional finite element analysis of translaminar lag screw designs for L5 bilateral spondylolysis.Frontiers in bioengineering and biotechnology · 2026
- Biomechanical comparison of laminoplasty and in-site regrafting versus laminectomy with pedicle screw fixation for localized thoracic pathological ossification of the ligamentum flavum: a finite element analysis.European spine journal : official publication of the European Spine Society, the European Spinal Deformity Society, and the European Section of the Cervical Spine Research Society · 2026
- Biomechanical comparative analysis of minimally invasive decompression and minimally invasive fusion with internal fixation in the treatment of lumbar spinal stenosis with degenerative spondylolisthesis: a finite element study.BMC musculoskeletal disorders · 2026
- Biomechanical study of posterior lumbar fixation combined with partial laminectomy using finite element analysisBIO Web of Conferences · 2026
- Biomechanical determinants of annular suture repair in an idealized post-discectomy annular defect: a finite element study.Frontiers in bioengineering and biotechnology · 2026
- Finite Element Comparison of Unilateral and Bilateral Polymethylmethacrylate Vertebral Fixation in the Canine Thoracolumbar Spine.Veterinary and comparative orthopaedics and traumatology : V.C.O.T · 2026