Biospecimen Integrity and Short-Term Analyte Stability after Pneumatic Tube Transport: Validation and Pressure Optimization of the TEMPUS 600 System in a Tertiary Care Hospital.
Dong-Won Yoo, Hyung-Hoi Kim
Department of Laboratory Medicine, Pusan National University Hospital, Busan, Republic of Korea. · Department of Laboratory Medicine, Biomedical Research Institute and Biomedical Informatics Unit, Pusan National University Hospital and Pusan National University School of Medicine, Busan, Republic of Korea.
Key points
From the paper's conclusion- Transport at 1.28 bar was associated with higher hemolysis index, LDH, and potassium than manual transport, and these elevations were no longer detected at 1.26 bar. Because the settings were evaluated sequentially and no equivalence margin was predefined, the findings support site-specific validation and operational optimization of PTS rather than a claim of equivalence.
Abstract
BACKGROUND: Pneumatic tube systems (PTS) reduce preanalytical delay but impose mechanical stress that can hemolyse specimens. We validated the TEMPUS 600 PTS at a tertiary hospital and sought an operating pressure that minimized transport-related hemolysis. METHODS: Specimens were those already collected for physician-ordered testing; the routine draw was divided between two smaller evacuated tubes of the same type, so no additional blood or urine was taken. In 80 patients, one 2.5 mL serum tube was hand-carried and one was sent by PTS at 1.28 bar; later the same day, the retained whole-blood residual of the hand-carried tube was sent at 1.26 bar, and five hemolysis-sensitive markers were compared by repeated-measures ANOVA. After 1.26 bar was implemented, PTS and manual transport were compared in 120 paired serum specimens (28 analytes) and 80 paired urine specimens (6 analytes), reported as bias following CLSI EP09. Turnaround time (TAT) was compared for the year before and after implementation. RESULTS: The hemolysis index differed across conditions (F[2,158] = 21.84, p < 0.001) PTS at 1.28 bar exceeded its contemporaneous manual control (10.20 vs. 6.22; p < 0.001), whereas at 1.26 bar, no difference from the manual baseline was detected (6.39 vs. 6.22; p = 0.79). Lactate dehydrogenase (LDH) and potassium behaved likewise. At 1.26 bar, bias was small and bidirectional (serum +6.36% to -14.64%; urine +3.90% to -6.94%). All three TAT components were shorter after implementation (p < 0.001). CONCLUSIONS: Transport at 1.28 bar was associated with higher hemolysis index, LDH, and potassium than manual transport, and these elevations were no longer detected at 1.26 bar. Because the settings were evaluated sequentially and no equivalence margin was predefined, the findings support site-specific validation and operational optimization of PTS rather than a claim of equivalence.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Performance evaluation of the pneumatic tube system for urine, haematology, coagulation, biochemical and immunological testsTurkish Journal of Biochemistry · 2026
- 251 Impact of BD MAP Microtube® Implementation on Hematology Turnaround TimeAmerican Journal of Clinical Pathology · 2026
- Pre-Analytical and Analytical Factors Affecting the Accuracy of Blood Gas Analysis—A Systematic Review with Structured Narrative SynthesisLaboratories · 2026
- 39 Reduction of Serum Hemolysis in the Emergency Department Using a Needleless PIVO™(BD) Blood Collection DeviceAmerican Journal of Clinical Pathology · 2026
- A-277 Improving Sample Integrity: Preclinical Evidence Demonstrating Reduced Hemolysis with Novel Blood Collection DeviceClinical Chemistry · 2026
- Effects of storage temperature and delayed analysis on routine biochemical and haematological test resultsScientific Reports · 2026