39 Reduction of Serum Hemolysis in the Emergency Department Using a Needleless PIVO™(BD) Blood Collection Device

Zoobia Khan, Yvette McCarter, Carmen Smotherman, Christine Gage, Nicole K Scotty, Zachary W Harrison, Katrina Agner, Matthew Feldhammer

Key points

From the paper's conclusion
  • Conclusion Implementation of the PIVO™(BD) device was associated with a large and statistically significant reduction in hemolysis compared with baseline practice and lower rates than concurrent standard collection methods.

Abstract

Background Hemolysis is a frequent pre-analytical error in emergency departments (EDs) that leads to specimen rejection, diagnostic delays, repeat phlebotomy, increased costs, and patient discomfort. Accurate results are particularly critical for high-sensitivity troponin and basic metabolic panel (BMP) testing. Blood collection through peripheral intravenous (IV) catheters is associated with higher hemolysis rates due to shear stress and negative pressure during aspiration. The PIVO™(BD) needleless blood collection device enables sampling through an existing peripheral IV while minimizing mechanical trauma to red blood cells. Objective To evaluate whether baseline implementation and use of the PIVO™(BD) device by trained ED staff reduces hemolysis rates compared with standard IV catheter collection methods and with non-PIVO™(BD) users. Methods We conducted a quality improvement initiative in a high-volume ED using a Plan-Do-Study-Act framework. Staff received standardized training on PIVO™(BD) collection. BMP and high-sensitivity troponin specimens obtained via PIVO™(BD) were prospectively labeled and tracked. Hemolysis was defined using laboratory hemolysis indices and specimen rejection data. Outcomes were compared among PIVO™(BD) users, PIVO™(BD) users not using the device during the same period (October 8–December 21, 2025), non-PIVO™(BD) users, and baseline data from three months earlier in the year (Jan 1–March 31, 2025). Results At baseline, 468 of 7,441 samples (6%) were hemolyzed prior to PIVO™(BD) implementation. Following adoption, 4 of 385 samples (1%) collected via PIVO™(BD) were hemolyzed, representing an 84% relative reduction (odds ratio = 0.16; 95% CI, 0.06–0.42; p < 0.001). During the implementation period, hemolysis occurred in 28 of 1,004 samples (2.79%) among PIVO™(BD) users not using the device and in 397 of 12,507 samples (3.17%) among non-PIVO™(BD) users. Historical data showed hemolysis in 50 of 2,212 samples (2.26%) among PIVO™(BD) users prior to consistent device use. Samples collected using PIVO™(BD) demonstrated consistently lower hemolysis rates than both baseline and concurrent standard collection methods. Conclusion Implementation of the PIVO™(BD) device was associated with a large and statistically significant reduction in hemolysis compared with baseline practice and lower rates than concurrent standard collection methods. By strengthening specimen integrity for time-sensitive testing, this scalable intervention has the potential to enhance diagnostic reliability, reduce repeat collections, and improve patient care in high-acuity ED settings. Adoption of PIVO™(BD) represents a practical, high-impact strategy to improve pre-analytical quality and laboratory performance.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers