American Journal of Clinical Pathology

251 Impact of BD MAP Microtube® Implementation on Hematology Turnaround Time

Bailee Thompson, Allison Hermane, Claudio Mosse, Joe Wiencek

Case report

Key points

From the paper's conclusion
  • We adopted it to automate small-volume processing and evaluated its 12-month impact.

Abstract

Introduction/Objective Standard microtainers® for small-volume hematology specimens cannot run on the automated line and require manual loading, repeats, and slide prep, adding turnaround time (TAT) and variability. The BD MAP Microtube® holds the same volume but has a false bottom compatible with laboratory automation. We adopted it to automate small-volume processing and evaluated its 12-month impact. Methods/Case Report We reviewed 89,050 hematology specimens at Monroe Carell Jr. Children’s Hospital at Vanderbilt (MCJCH), where short draws (or small volume) are most concentrated, over 20 months (September 2024 to April 2026). Adoption grew from 0% to 90% over the study period. Differences were tested with Mann-Whitney U and Brown-Forsythe. Monthly TAT was regressed on adoption. Statistical Process Control (SPC) and P-charts tracked trends and 60/120-minute compliance. Results Mean TAT dropped from 30.9 to 26.7 min. Each percentage point increase in adoption reduced TAT by 0.047 minutes (R² = 0.51, p Conclusion The implementation of the BD MAP Microtube® significantly decreased mean TAT and reduced TAT variability at Vanderbilt Medical Laboratories.

Read the paper