Evaluation of Point-of-Care Testing for Creatinine, Estimated Glomerular Filtration Rate, and Potassium in a Diverse United Kingdom Population.

Rouvick Mariano Gama, Amy Packham-Smith, Abigail Clark, Emmanuel Joson, Nirmaladevi Santhosh, Danilo Nebres, Kathryn Dalrymple, Royce P Vincent,

Faculty of Life Sciences and Medicine, King's College London, London, United Kingdom. · King's Kidney Care, King's College Hospital, London, United Kingdom.

Observational · n=310

Key points

From the paper's conclusion
  • Epoc demonstrated superior overall analytical agreement with laboratory SCr and performance for CKD classification, compared with NovaMax Pro. Device-specific biases and potassium variability highlight the importance of sampling technique, clinical context, and intended use when implementing renal POCT.

Abstract

BACKGROUND: Point-of-care testing (POCT) for renal function has the potential to enhance access to kidney disease management. However, real-world validation of renal POCT devices in people with chronic kidney disease (CKD) remains limited, particularly in diverse populations. METHODS: We conducted a prospective cohort study to evaluate agreement between 2 renal POCT devices for serum creatinine (SCr), estimated glomerular filtration rate (eGFR), and potassium. Adults with and without CKD underwent capillary and venous sampling on POCT devices, with comparison to reference laboratory standards. Creatinine was measured using NovaMax Pro and Epoc Blood Analysis System (Epoc); potassium was assessed with Epoc. Agreement was evaluated using correlation coefficients, bias, precision, accuracy, and Bland-Altman analysis. Sensitivity and specificity for detecting CKD (eGFR < 60 mL/min/1.73 m2) were evaluated. RESULTS: There were 310 participants included; 49% were from ethnic minority groups (29% Black; 14% South Asian). Both POCT devices demonstrated strong correlation with laboratory SCr and eGFR. NovaMax Pro systematically overestimated SCr, particularly at higher concentrations, whereas Epoc showed greater precision and accuracy. No significant differences were observed between capillary and venous samples. eGFR bias was modest for both devices. Sensitivity was higher with NovaMax Pro (94%), while specificity was higher with Epoc (99%). Epoc potassium measurements differed significantly between capillary and venous samples (P < 0.001). CONCLUSION: Epoc demonstrated superior overall analytical agreement with laboratory SCr and performance for CKD classification, compared with NovaMax Pro. Device-specific biases and potassium variability highlight the importance of sampling technique, clinical context, and intended use when implementing renal POCT.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers